FDA Adverse Event Malfunction Summary report: N

GE OEC 9800

MDR report key: 1071150 · Received July 10, 2008

Report

Report Number
1720753-2008-23091
Event Type
Malfunction
Date Received
July 10, 2008
Date of Event
June 5, 2008
Report Date
July 1, 2008
Manufacturer
GE OEC MEDICAL SYSTEMS INC.
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REP PERFORMED AN ON SITE INVESTIGATION. THE PS2 POWER SUPPLY FUSE WAS REPLACED AND THE POWER SUPPLY WAS ADJUSTED DURING THE SERVICE CALL. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED, AND PUT BACK INTO SERVICE.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE 9800 SYSTEM HAD AN INTERLOCK FAILURE ERROR MESSAGE AND THE VERTICAL LIFT COLUMN WOULD NOT WORK PRIOR TO A CASE. NO PT INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 GE OEC 9800 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS INC. NA NA

Patients

Seq Age Sex Outcome Treatment
1