FDA Adverse Event Malfunction Summary report: N

GE OEC 9800

MDR report key: 1071061 · Received July 10, 2008

Report

Report Number
1720753-2008-22868
Event Type
Malfunction
Date Received
July 10, 2008
Date of Event
March 4, 2008
Report Date
June 25, 2008
Manufacturer
GE OEC MEDICAL SYSTEMS, INC.
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MD, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REP PERFORMED AN ON SITE INVESTIGATION. THE IMAGE PROCESSOR WAS REPLACED AND THE DISPLAY ADAPTOR WAS RE-SEATED DURING THE SERVICE CALL. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED, AND PUT BACK INTO SERVICE.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE 9800 SYSTEM HAD HORIZONTAL LINES AND DARK IMAGES. NO PT INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 GE OEC 9800 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS, INC. NA NA

Patients

Seq Age Sex Outcome Treatment
1