PUMP MMT-1781K 670G V4.11 MG
Report
- Report Number
- 2032227-2020-194099
- Event Type
- Malfunction
- Date Received
- October 21, 2020
- Date of Event
- July 18, 2020
- Report Date
- May 3, 2022
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OZP
- UDI-DI
- 00763000187415
- Removal / Correction Number
- Z-0956-2020
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SP
- Reporter Occupation
- OTHER
- Health Professional
- N
Narratives
THIS REPORT IS PART OF A RETROSPECTIVE REVIEW AND REMEDIATION EFFORTS IN RESPONSE TO A WARNING LETTER. UPDATED H9: Z-0956-2020. MEDTRONIC, INC. (MEDTRONIC) IS SUBMITTING THIS REPORT TO COMPLY WITH 21 C.F.R. PART 803, THE MEDICAL DEVICE REPORTING REGULATION. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION IN THE TIME ALLOTTED AND HAS PROVIDED AS MUCH INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY "DEFECTS" OR HAS "MALFUNCTIONED". THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA, TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REPORTING PURSUANT TO PART 803. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE IT BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT.
(B)(4). UNIT RECEIVED WITH BLANK DISPLAY ANOMALY DUE TO MOISTURE DAMAGE ON ELECTRONICS ASSEMBLY. UNABLE TO PERFORMED DISPLACEMENT, REWIND, SEATING AND BASIC OCCLUSION DUE TO BLANK DISPLAY ANOMALY. UNIT RECEIVED WITH MOISTURE DAMAGE NOTED ON MOTOR, VIBRATOR MOTOR OR BATTERY TUBE ASSEMBLY. UNIT RECEIVED WITH CRACKED RETAINER, CRACKED BATTERY TUBE THREADS, FADING SERIAL NUMBER LABEL AND CRACKED CASE AT BATTERY TUBE SIDE. CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT INSULIN PUMP WAS EXPOSED TO BEACH WATER. CUSTOMER WENT TO THE BEACH AND THERE WAS WATER UNDER LCD DISPLAY. CUSTOMER STATED THE INSULIN PUMP WAS NOT DROPPED. CUSTOMER STATED THERE ARE NOT CRACKS IN THE INSULIN PUMP. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE CUSTOMER WAS ASSISTED WITH TROUBLESHOOTING. THE DEVICE WILL BE RETURNED FOR ANALYSIS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1179647 | PUMP MMT-1781K 670G V4.11 MG | ARTIFICIAL PANCREAS DEVICE SYSTEM, SINGLE HORMONAL CONTROL | OZP | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-1781K | HG3GVE3 | 00763000187415 |