8300 ALARIS ETCO2 MODULE
Report
- Report Number
- 2016493-2020-24858
- Event Type
- Malfunction
- Date Received
- October 21, 2020
- Report Date
- March 13, 2018
- Manufacturer
- CAREFUSION SD
- Product Code
- CCK
- PMA / PMN Number
- K031741
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- 003
Narratives
THIS REPORTED EVENT AND SUBSEQUENT REPAIRS WERE INVESTIGATED THROUGH THE TSC TROUBLESHOOTING PROCESS. A REVIEW OF THE DEVICE SERVICE HISTORY RECORD WAS PERFORMED FROM THE DATE OF MANUFACTURE TO THE DATE CORRESPONDING TO THIS SERVICE NOTIFICATION NUMBER. THE DEVICE WAS NOT PREVIOUSLY RETURNED FOR SERVICE. THE DATABASE SHOWED NO QUALITY NOTIFICATIONS WERE OPENED FOR THE DEVICE. A DHR (DEVICE HISTORY RECORD) REVIEW CANNOT BE COMPLETED AS THE SERIAL NUMBER WAS NOT OBTAINED UPON RECEIPT OF THE COMPLAINT. ADDITIONALLY, A HISTORICAL REVIEW OF COMPLAINTS IN TRACKWISE CANNOT BE CONDUCTED. CAPA REFERENCE: (B)(4). THE CUSTOMER STATED THAT THERE WAS NO PATIENT INVOLVEMENT.
CASE ID: (B)(4). CASE STATUS: OPEN. SUMMARY: SENSOR CALIBRATION TEST. CASE NOTES: (B)(4). NPI. ETCO2 UNIT. CUSTOMER (B)(6) CALLED IN FOR HELP ON ETCO2 UNIT. HE IS TRY TO RUN SENSOR CALIBRATION WITH THE GAS CONNECT AND FLOW METER, BUT UNIT IS FAILING ON SAME SETUP THREE UNITS SO FAR FAILING. THEY RECEIVE CASE OF THE GAS THINKS HE HAS SOME BAD CANS HE WILL CALL THE COMPANY HE ORDER THE GAS FROM AIR LIQUID AND GET THOSE REPLACE CALL US BACK IF HE NEEDS TO.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1178484 | 8300 ALARIS ETCO2 MODULE | PUMP, INFUSION | CCK | CAREFUSION SD | 8300 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |