FDA Adverse Event Malfunction Summary report: N

8300 ALARIS ETCO2 MODULE

MDR report key: 10710086 · Received October 21, 2020

Report

Report Number
2016493-2020-24858
Event Type
Malfunction
Date Received
October 21, 2020
Report Date
March 13, 2018
Manufacturer
CAREFUSION SD
Product Code
CCK
PMA / PMN Number
K031741
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 1

THIS REPORTED EVENT AND SUBSEQUENT REPAIRS WERE INVESTIGATED THROUGH THE TSC TROUBLESHOOTING PROCESS. A REVIEW OF THE DEVICE SERVICE HISTORY RECORD WAS PERFORMED FROM THE DATE OF MANUFACTURE TO THE DATE CORRESPONDING TO THIS SERVICE NOTIFICATION NUMBER. THE DEVICE WAS NOT PREVIOUSLY RETURNED FOR SERVICE. THE DATABASE SHOWED NO QUALITY NOTIFICATIONS WERE OPENED FOR THE DEVICE. A DHR (DEVICE HISTORY RECORD) REVIEW CANNOT BE COMPLETED AS THE SERIAL NUMBER WAS NOT OBTAINED UPON RECEIPT OF THE COMPLAINT. ADDITIONALLY, A HISTORICAL REVIEW OF COMPLAINTS IN TRACKWISE CANNOT BE CONDUCTED. CAPA REFERENCE: (B)(4). THE CUSTOMER STATED THAT THERE WAS NO PATIENT INVOLVEMENT.

Description of Event or Problem · 1

CASE ID: (B)(4). CASE STATUS: OPEN. SUMMARY: SENSOR CALIBRATION TEST. CASE NOTES: (B)(4). NPI. ETCO2 UNIT. CUSTOMER (B)(6) CALLED IN FOR HELP ON ETCO2 UNIT. HE IS TRY TO RUN SENSOR CALIBRATION WITH THE GAS CONNECT AND FLOW METER, BUT UNIT IS FAILING ON SAME SETUP THREE UNITS SO FAR FAILING. THEY RECEIVE CASE OF THE GAS THINKS HE HAS SOME BAD CANS HE WILL CALL THE COMPANY HE ORDER THE GAS FROM AIR LIQUID AND GET THOSE REPLACE CALL US BACK IF HE NEEDS TO.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1178484 8300 ALARIS ETCO2 MODULE PUMP, INFUSION CCK CAREFUSION SD 8300

Patients

Seq Age Sex Outcome Treatment
1