NEXGEN COMPLETE KNEE SOLUTION MIS TOTAL KNEE PROCEDURE TIBIAL BROACH IMPACTOR
Report
- Report Number
- 1822565-2008-00394
- Event Type
- Malfunction
- Date Received
- July 3, 2008
- Date of Event
- April 29, 2008
- Report Date
- May 5, 2008
- Manufacturer
- ZIMMER, INC.
- Product Code
- HWA
- Removal / Correction Number
- 1822565-07/01/2008-002-R
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- PHYSICIAN
Narratives
EVALUATION SUMMARY: THE RETURNED BROACH IMPACTOR HAS ONE OF THE JAWS FRACTURED OFF FROM THE BASE. THE KNURLED SECTION OF THE INSTRUMENT SHOWS IMPACTION MARKS ON THE ANTERIOR SIDE INDICATING THAT THE DEVICE WAS IMPACTED OFF-AXIS CAUSING ADDITIONAL BENDING MOMENT AND THE JAW FRACTURED DUE TO OVERLOAD. IMPROPER USE OF THE DEVICE APPEARS TO BE THE CAUSE OF THE PROBLEM. A FIELD COMMUNICATION WAS RELEASED JANUARY 2008 FOR PROPER EXTRACTION TECHNIQUE OF THE BROACH IMPACTOR. IN ADDITION, A NEW DESIGN HAS BEEN IMPLEMENTED THAT CAN SUCCESSFULLY WITHSTAND OFF-AXIS LOADING. A PRODUCT REMOVAL OF ALL PREVIOUS DESIGN ITERATIONS WAS INITIATED ON 06/30/08. EVALUATION: THE FRACTURED TAB WAS NOT RETURNED WITH THE COMPLAINT FOR REVIEW. THE DEVICE MEETS SPECIFICATIONS AS MEASURED.
IT IS REPORTED THAT BROACH IMPACTOR IS BROKEN.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | NEXGEN COMPLETE KNEE SOLUTION MIS TOTAL KNEE PROCEDURE TIBIAL BROACH IMPACTOR | KNEE INSTRUMENT | HWA | ZIMMER, INC. | NA | 60356560 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK |