FDA Adverse Event
Malfunction
Summary report: N
THERMOPHORE AUTOMATIC MOIST HEAT PACK
MDR report key: 1070948
·
Received June 30, 2008
Report
- Report Number
- 1811605-2008-00100
- Event Type
- Malfunction
- Date Received
- June 30, 2008
- Date of Event
- June 1, 2008
- Report Date
- June 18, 2008
- Manufacturer
- BATTLE CREEK EQUIPMENT CO.
- Product Code
- IRT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
Description of Event or Problem · 1
IT WAS REPORTED THE ELECTRICAL CORD EMITTED SPARKS. VISUAL INSPECTION OF THE UNIT REVEALED A BREAK IN THE LINE CORD NEAR THE STRAIN RELIEF.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | THERMOPHORE AUTOMATIC MOIST HEAT PACK | ELECTRIC MOIST HEAT PACK | IRT | BATTLE CREEK EQUIPMENT CO. | 055 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |