FDA Adverse Event
Injury
Summary report: N
ANIMAS INSULIN INFUSION PUMP
MDR report key: 1070917
·
Received July 7, 2008
Report
- Report Number
- 2531779-2008-00378
- Event Type
- Injury
- Date Received
- July 7, 2008
- Date of Event
- June 9, 2008
- Report Date
- June 9, 2008
- Manufacturer
- ANIMAS CORP.
- Product Code
- LZG
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NC, US
- Reporter Occupation
- UNKNOWN
Narratives
Additional Manufacturer Narrative · 1
THE PUMP WAS RETURNED TO ANIMAS FOR EVAL. EVAL REVEALED A VISIBLE BATTERY COMPARTMENT CRACK AND A BLANK DISPLAY SCREEN, BUT DEMONSTRATED THAT PUMP INSULIN DELIVERY WAS OPERATING WITHIN REQUIRED SPECS AND DELIVERY ACCURACY.
Description of Event or Problem · 1
THE PT WAS HOSPITALIZED FOR ELEVATED BLOOD GLUCOSE LEVELS. THE PT'S HUSBAND ALSO REPORTED THAT THE PUMP EXHIBITED A DAMAGED DISPLAY SCREEN.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ANIMAS INSULIN INFUSION PUMP | INSULIN INFUSION PUMP | LZG | ANIMAS CORP. | IR 1200 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Hospitalization |