FDA Adverse Event Malfunction Summary report: N

MENTOR CPX 4 BREAST TISSUE EXPANDER WITH SUTURE TABS

MDR report key: 10709018 · Received October 20, 2020

Report

Report Number
1645337-2020-13358
Event Type
Malfunction
Date Received
October 20, 2020
Date of Event
January 4, 2020
Report Date
September 30, 2020
Manufacturer
MENTOR TEXAS
Product Code
LCJ
PMA / PMN Number
K130813
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
BR
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 0

ON 31-MAR-2021, MENTOR RECEIVED ADDITIONAL INFORMATION ABOUT THE EVENT. THE SUSPECT MEDICAL DEVICE IS A MENTOR CPX 4 BREAST TISSUE EXPANDER WITH SUTURE TABS 350CC DEVICE, CATALOG #3549212, LOT #7663383, UDI #(B)(4), 510(K) #K130813. A MANUFACTURING RECORD EVALUATION (MRE) WAS PERFORMED, AND NO ANOMALIES WERE FOUND RELATED TO THIS COMPLAINT. IN ADDITION, THE MRE VERIFIES THAT THE DEVICE WAS MANUFACTURED IN ACCORDANCE WITH DOCUMENTED SPECIFICATION AND PROCEDURES. ON 12-APR-2021, THE MENTOR FAILURE ANALYSIS LAB RECEIVED THE DEVICE FOR EVALUATION. THE ANALYSIS HAS BEGUN BUT IS NOT COMPLETE AT THIS TIME. WHEN THE INVESTIGATIONAL ANALYSIS HAS BEEN COMPLETED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. MANUFACTURER¿S REFERENCE NUMBER: (B)(4).

Additional Manufacturer Narrative · 0

ON 04-MAY-2021, MENTOR COMPLETED THE INVESTIGATION ON THE SUSPECT MEDICAL DEVICE. DEVICE EVALUATION SUMMARY: ACCORDING TO THE INFORMATION RECEIVED, THE PATIENT EXPERIENCED A DEFLATION IN THE TISSUE EXPANDER DURING SURGERY. THE PRODUCT WAS RETURNED TO MENTOR FOR EVALUATION. MENTOR CONDUCTED A VISUAL INSPECTION LEAK TEST AND MICROSCOPIC EXAMINATION OF THE RETURNED DEVICE. VISUAL ANALYSIS OF THE RETURNED SAMPLE REVEALED THAT THE TISSUE EXPANDER WAS FOUND INTACT. LEAK TESTING WAS PERFORMED, IN ACCORDANCE WITH MENTOR PROCEDURES AND A TEAR WAS NOTED ON ANTERIOR ASPECT MEASURING APPROXIMATELY 0.1 CM. MICROSCOPIC EXAMINATION WAS PERFORMED AND THE CAUSE OF THE TEAR COULD NOT BE IDENTIFIED. A MANUFACTURING RECORD EVALUATION WAS PERFORMED FOR THE FINISHED DEVICE BATCH NUMBER, AND NO NON-CONFORMANCES WERE IDENTIFIED. AS PART OF MENTOR¿S QUALITY PROCESS ALL DEVICES ARE MANUFACTURED, INSPECTED, AND RELEASED TO APPROVED SPECIFICATIONS. ADDITIONAL COMPLAINT INFORMATION MONITORING FOR POTENTIAL SAFETY SIGNALS IS CONDUCTED THROUGH COMPLAINT TRENDING AS PART OF POST MARKET SURVEILLANCE. MANUFACTURER¿S REFERENCE NUMBER: (B)(4).

Additional Manufacturer Narrative · 1

SINCE THE DEVICE HAS NOT BEEN RETURNED FOR ANALYSIS, NO PRODUCT FAILURE ANALYSIS CAN BE CONDUCTED, AND NO DETERMINATION OF POSSIBLE CONTRIBUTING FACTORS CAN BE MADE. AS SUCH, THE INVESTIGATION WILL BE CLOSED. IF THE COMPLAINT DEVICE IS RECEIVED IN THE FUTURE, THE INVESTIGATION WILL BE REOPENED AND CONDUCTED AS APPROPRIATE. SINCE NO LOT NUMBER WAS PROVIDED, NO MANUFACTURING RECORD EVALUATION COULD BE PERFORMED. MANUFACTURER¿S REFERENCE NUMBER: (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT AN UNSPECIFIED MENTOR TISSUE EXPANDER DEVICE DEFLATED DURING A SCHEDULED BREAST SURGERY. THERE WAS NO REPORTED CONSEQUENCE TO THE PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1167534 MENTOR CPX 4 BREAST TISSUE EXPANDER WITH SUTURE TABS EXPANDER, SKIN, INFLATABLE LCJ MENTOR TEXAS 7663383

Patients

Seq Age Sex Outcome Treatment
1