FDA Adverse Event Malfunction Summary report: N

9900

MDR report key: 1070871 · Received July 9, 2008

Report

Report Number
1720753-2008-22835
Event Type
Malfunction
Date Received
July 9, 2008
Date of Event
June 12, 2008
Report Date
June 24, 2008
Manufacturer
GE OEC MEDICAL SYSTEMS INC.
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THE GE SERVICE REP TORQUED WORKSTATION TRANSFORMER NUT'S TO 60" LBS CHECKED ALL WORKSTATION FUSE AND FUSE HOLDER SWITCH SERVICE UP TO VERIFIED NO ERRORS ON BOOT-UP. BOOTED SYSTEM X 10 SYSTEM BOOTS WITHIN 165 SEC. SYSTEM CHECKED AND TESTED O.K.

Description of Event or Problem · 1

THE CUSTOMER REPORTED INTERMITTENT SLOW TO BOOT UP. NO PT INVOLVED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 9900 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS INC. NA NA

Patients

Seq Age Sex Outcome Treatment
1