GPS PROCEDURAL STRETCHER
MDR report key: 1070833
·
Received July 3, 2008
Report
- Report Number
- 1824206-2008-02790
- Event Type
- Malfunction
- Date Received
- July 3, 2008
- Date of Event
- June 30, 2008
- Report Date
- June 30, 2008
- Manufacturer
- HILL-ROM RITTER
- Product Code
- FPO
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NC, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE BRAKES NOT HOLDING/WORKING COULD LEAD TO UNINTENTIONAL MOVEMENT OF THE STRETCHER WHICH HAS THE POTENTIAL TO CAUSE SERIOUS INJURY. THE TECHNICIAN REPLACED WORN CASTERS IN ORDER TO RESOLVE THE PROBLEM.
Description of Event or Problem · 1
ACCOUNT STATED THAT BRAKES WOULD NOT HOLD WHEN THE BRAKES WERE APPLIED.