FDA Adverse Event Malfunction Summary report: N

MICROSPEED UNI CONTROL UNIT W/COOL.UNIT

MDR report key: 1070820 · Received July 3, 2008

Report

Report Number
2916714-2008-00012
Event Type
Malfunction
Date Received
July 3, 2008
Report Date
June 13, 2008
Manufacturer
AESCULAP AG & CO. KG
Product Code
HBC
PMA / PMN Number
K053526
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
RISK MANAGER

Narratives

Additional Manufacturer Narrative · 1

REPORT INCLUDES COMPONENT WHICH WAS BEING USED WITH THE CONTROL UNIT: MODEL#: GD455M, BRAND NAME: MICRO-LINE ANGULAR HDPC 1:2 F/2.35X70MM, COMMON DEVICE NAME: HWE, FACILITY REPORTED THE KLS MARTIN DRILL GUIDE WAS SUPPOSED TO BE RUN ON 5,000 RPM'S OR LESS AND WAS RUN ON 30,000 RPM'S. AESCULAP, INC ANTICIPATES THE DEVICE WILL BE RELEASED FROM THE USER FACILITY FOR EVAL; HOWEVER, IT IS UNKNOWN IF IT WILL BE RELEASED.

Description of Event or Problem · 1

WHILE USING THE MOTOR, HANDPIECE, AND A KLS MARTIN DRILL GUIDE ON AN OMF CASE. PATIENT WAS BURNED, ON THEIR CHEEK, WHILE IN OPERATION. INJURY WAS MINOR AND WAS TREATED WITH TOPICAL OINTMENT. SURGERY WAS NOT DELAYED. INJURY WAS SUPERFICIAL AND CONSISTED OF A SMALL BLISTER ON THE FACE NEAR INCISION.