DIMENSION CLINICAL CHEMISTRY SYSTEM
Report
- Report Number
- 2517506-2008-00055
- Event Type
- Other
- Date Received
- June 27, 2008
- Date of Event
- June 9, 2008
- Report Date
- June 9, 2008
- Manufacturer
- SIEMENS HEALTHCARE DIAGNOSTICS, INC.
- Product Code
- MMI
- PMA / PMN Number
- K010313
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MI, US
- Reporter Occupation
- OTHER
Narratives
ANALYSIS OF THE INSTRUMENT AND INSTRUMENT DATA INDICATE THAT THE CAUSE FOR THE FALSELY ELEVATED TROPONIN I RESULT WAS SAMPLE INTEGRITY. THE IFU FOR DIMENSION CTNI FLEX REAGENT CARTRIDGE CONTAINS THE FOLLOWING INFO: "SPECIMENS SHOULD BE FREE OF PARTICULATE MATTER TO PREVENT APPEARANCE OF FIBRIN IN SERUM SAMPLES, COMPLETE CLOT FORMATION SHOULD TAKE PLACE BEFORE CENTRIFUGATION." THE INSTRUMENT IS PERFORMING WITHIN SPECIFICATIONS. NO FURTHER EVALUATION OF THE DEVICE IS REQUIRED.
A FALSELY ELEVATED TROPONIN I RESULT WAS OBTAINED ON A PT SAMPLE. THE RESULT WAS NOT REPORTED TO THE PHYSICIAN. THE SAMPLE WAS REPEATED ON AN ALTERNATE ANALYZER AND A NEGATIVE RESULT WAS OBTAINED. PATIENT TREATMENT WAS NOT ALTERED OR PRESCRIBED. THERE WAS NO REPORT OF ADVERSE HEALTH CONSEQUENCES AS A RESULT OF THE FALSELY ELEVATED TROPONIN I RESULT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | DIMENSION CLINICAL CHEMISTRY SYSTEM | CTNI FLEX REAGENT CARTRIDGE | MMI | SIEMENS HEALTHCARE DIAGNOSTICS, INC. | NA | FA9076 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |