FDA Adverse Event Malfunction Summary report: N

LIFEPAK 20 DEFIBRILLATOR/MONITOR

MDR report key: 1070692 · Received July 2, 2008

Report

Report Number
3015876-2008-00742
Event Type
Malfunction
Date Received
July 2, 2008
Date of Event
June 6, 2008
Report Date
June 6, 2008
Manufacturer
PHYSIO-CONTROL, INC.
Product Code
LDD
PMA / PMN Number
K012274
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MA, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

PHYSIO-CONTROL EVALUATED THE DEVICE AND VERIFIED THE REPORTED FAILURE. PHYSIO REPLACED THE THERAPY BOARD PCB ASSEMBLY AND THEN OBSERVED PROPER DEVICE OPERATION THROUGH FUNCTIONAL AND PERFORMANCE TESTING. THE DEVICE WAS RETURNED TO THE CUSTOMER FOR USE. PHYSIO FURTHER EVALUATED THE REPLACED ASSEMBLY AND DETERMINED THAT ROOT CAUSE OF THE REPORTED FAILURE WAS DUE TO IC CHIP, DESIGNATOR U61.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE DEVICE WAS GIVING "ABNORMAL ENERGY DELIVERY" WHEN SHOCKING INTO A SIMULATOR. THERE WAS NO PATIENT USE ASSOCIATED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 LIFEPAK 20 DEFIBRILLATOR/MONITOR LDD PHYSIO-CONTROL, INC. 20 NA

Patients

Seq Age Sex Outcome Treatment
1 NA