DEXCOM G6 CONTINUOUS GLUCOSE MONITORING SYSTEM
MDR report key: 10706561
·
Received October 20, 2020
Report
- Report Number
- 3004753838-2020-122577
- Event Type
- Malfunction
- Date Received
- October 20, 2020
- Date of Event
- September 30, 2020
- Report Date
- April 19, 2021
- Manufacturer
- DEXCOM, INC.
- Product Code
- QBJ
- UDI-DI
- 00386270000385
- PMA / PMN Number
- DEN170088
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NH, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 0
COM-2012075-20
Additional Manufacturer Narrative · 0
COM-2012075-20.
Description of Event or Problem · 0
IT WAS REPORTED THAT TRANSMITTER FAILED ERROR OCCURRED. DATA WAS EVALUATED AND THE ALLEGATION WAS CONFIRMED. THE PROBABLE CAUSE WAS DETERMINED TO BE A LOW TRANSMITTER BATTERY. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Description of Event or Problem · 0
IT WAS REPORTED THAT TRANSMITTER FAILED ERROR OCCURRED. THE PRODUCT WAS EVALUATED. AN EXTERNAL VISUAL INSPECTION WAS PERFORMED AND PASSED. VOLTAGE TEST WAS PERFORMED AND FAILED DUE TO 0V. A REVIEW OF THE SHARE LOGS WAS PERFORMED AND TRANSMITTER FAILED ERROR WAS FOUND WITHIN THE INVESTIGATION WINDOW. THE ALLEGATION WAS CONFIRMED. THE PROBABLE CAUSE WAS DETERMINED TO BE A LOW TRANSMITTER BATTERY. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.