FDA Adverse Event Malfunction Summary report: N

CONTOUR TEST STRIPS (50)

MDR report key: 1070614 · Received July 1, 2008

Report

Report Number
1826988-2008-00729
Event Type
Malfunction
Date Received
July 1, 2008
Date of Event
June 16, 2008
Report Date
June 16, 2008
Manufacturer
BAYER HEALTHCARE, LLC
Product Code
NBW
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
VA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE QA LAB FOUND THE RETURNED REAGENT TO READ 2MG/DL HIGH, OUT OF SPECIFICATION. PERFORMANCE WAS SATISFACTORY USING IQA RETENTION REAGENT.

Description of Event or Problem · 1

THE CUSTOMER STATED THAT SHE RECEIVED A HIGH GLUCOSE READING OF 446 MG/DL. WHILE TROUBLESHOOTING, SHE PERFORMED CONTROL TESTS AND RECEIVED A RESULT OF 288 MG/DL. THE NORMAL CONTROL RANGE WAS 103-142 MG/DL. NO ADVERSE EVENTS WERE ALLEGED. THE TEST STRIPS WERE RETURNED FOR EVALUATION. REPLACEMENT TEST STRIPS WERE SENT TO THE CUSTOMER.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CONTOUR TEST STRIPS (50) BLOOD GLUCOSE TEST STRIPS NBW BAYER HEALTHCARE, LLC 7KC3D09

Patients

Seq Age Sex Outcome Treatment
1 UNK