FDA Adverse Event Malfunction Summary report: N

HARMONIC ACE 36 CM

MDR report key: 1070585 · Received July 2, 2008

Report

Report Number
3005075853-2008-00255
Event Type
Malfunction
Date Received
July 2, 2008
Date of Event
May 23, 2008
Report Date
May 25, 2008
Manufacturer
ETHICON ENDO-SURGERY, LLC
Product Code
LFL
PMA / PMN Number
K042777
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

DATE SENT: 7/2/2008. EVAL SUMMARY: THE DEVICE WAS RECEIVED IN GOOD CONDITION. NO VISIBLE DAMAGE WAS NOTED. THE HAND-ACTIVATION CONTACTS WERE IN GOOD CONDITION. THE DEVICE WAS TESTED ON A GENERATOR AND IT WAS FOUND THAT THE HAND CONTROL SWITCH ASSEMBLY BUTTONS WERE NOT FUNCTIONAL. HOWEVER, WITH FOOT SWITCH ASSEMBLY IT WORKED AS INTENDED. COMPLAINT INFO IS TRENDED ON A REGULAR BASIS TO DETERMINE IF FURTHER INVESTIGATION IS WARRANTED. THE BATCH HISTORY RECORDS WERE REVIEWED WITH NO ANOMALIES NOTED DURING THE MANUFACTURING PROCESS.

Description of Event or Problem · 1

IT WAS REPORTED DURING AN UNK PROCEDURE THAT THE HAND SWITCH ERROR DISPLAYED AND COULD NOT ACTIVATE. ANOTHER LIKE DEVICE WAS USED. THERE WAS NO PT CONSEQUENCE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 HARMONIC ACE 36 CM LFL ETHICON ENDO-SURGERY, LLC NA E4KL1T

Patients

Seq Age Sex Outcome Treatment
1 GENERATOR| HANDPIECE