FDA Adverse Event
Malfunction
Summary report: N
HARMONIC ACE 36 CM
MDR report key: 1070585
·
Received July 2, 2008
Report
- Report Number
- 3005075853-2008-00255
- Event Type
- Malfunction
- Date Received
- July 2, 2008
- Date of Event
- May 23, 2008
- Report Date
- May 25, 2008
- Manufacturer
- ETHICON ENDO-SURGERY, LLC
- Product Code
- LFL
- PMA / PMN Number
- K042777
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
DATE SENT: 7/2/2008. EVAL SUMMARY: THE DEVICE WAS RECEIVED IN GOOD CONDITION. NO VISIBLE DAMAGE WAS NOTED. THE HAND-ACTIVATION CONTACTS WERE IN GOOD CONDITION. THE DEVICE WAS TESTED ON A GENERATOR AND IT WAS FOUND THAT THE HAND CONTROL SWITCH ASSEMBLY BUTTONS WERE NOT FUNCTIONAL. HOWEVER, WITH FOOT SWITCH ASSEMBLY IT WORKED AS INTENDED. COMPLAINT INFO IS TRENDED ON A REGULAR BASIS TO DETERMINE IF FURTHER INVESTIGATION IS WARRANTED. THE BATCH HISTORY RECORDS WERE REVIEWED WITH NO ANOMALIES NOTED DURING THE MANUFACTURING PROCESS.
Description of Event or Problem · 1
IT WAS REPORTED DURING AN UNK PROCEDURE THAT THE HAND SWITCH ERROR DISPLAYED AND COULD NOT ACTIVATE. ANOTHER LIKE DEVICE WAS USED. THERE WAS NO PT CONSEQUENCE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | HARMONIC ACE 36 CM | LFL | ETHICON ENDO-SURGERY, LLC | NA | E4KL1T |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | GENERATOR| HANDPIECE |