FDA Adverse Event Malfunction Summary report: N

SYRINGE NON STERILE 50ML LL

MDR report key: 10705115 · Received October 20, 2020

Report

Report Number
3003152976-2020-00455
Event Type
Malfunction
Date Received
October 20, 2020
Date of Event
September 25, 2020
Report Date
October 21, 2020
Manufacturer
BECTON DICKINSON, S.A.
Product Code
FMF
PMA / PMN Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IT
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

H.6. INVESTIGATION: THREE PHOTOS WERE PROVIDED TO OUR QUALITY TEAM FOR INVESTIGATION. THROUGH VISUAL INSPECTION, A LEAKAGE PAST THE STOPPER RIBS WAS OBSERVED. BASED ON THE PHOTO, THE STOPPER IS ASSEMBLED PROPERLY ONTO THE PLUNGER AND THERE IS NO VISIBLE DAMAGE OR DEFECTS THAT COULD HAVE CONTRIBUTED TO THE REPORTED LEAK. A DEVICE HISTORY REVIEW WAS PERFORMED FOR THE REPORTED LOT 1904352, NO DEVIATIONS OR NON-CONFORMANCES WERE IDENTIFIED DURING THE MANUFACTURING PROCESS THAT COULD HAVE CONTRIBUTED TO THIS ISSUE. FINAL PRODUCTS IN THIS MANUFACTURING LINE, FOR THIS REFERENCE ARE SAMPLED AND THEY ARE SUBJECTED TO VISUAL AND FUNCTIONAL INSPECTIONS DURING THE DIFFERENT MANUFACTURING SUB-PROCESSES ACCORDING TO PROCEDURES, INCLUDING TIGHTNESS VERIFICATION OF THE STOPPER. PRODUCT THAT FAILS THIS TESTING IS AUTOMATICALLY REJECTED. BASED ON THE AVAILABLE INFORMATION WE ARE NOT ABLE TO DETERMINE A ROOT CAUSE AT THIS TIME. H3 OTHER TEXT : SEE H.10.

Description of Event or Problem · 0

IT WAS REPORTED THAT SYRINGE NON STERILE 50ML LL LEAKED BEFORE USE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: I AM WRITING TO YOU BECAUSE THIS MORNING WE RECEIVED A REPORT FROM ONE OF OUR CUSTOMERS ABOUT A DEFECTIVE ASN-50 SYRINGE. IN PARTICULAR THE DEFECT IS THE LEAKAGE OF LIQUIDS FROM THE PISTON OF THE SYRINGE, AS SHOWN IN THE PICTURES.

Additional Manufacturer Narrative · 1

A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED, BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT SYRINGE NON STERILE 50 ML LL LEAKED BEFORE USE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: I AM WRITING TO YOU BECAUSE THIS MORNING WE RECEIVED A REPORT FROM ONE OF OUR CUSTOMERS ABOUT A DEFECTIVE ASN-50 SYRINGE. IN PARTICULAR THE DEFECT IS THE LEAKAGE OF LIQUIDS FROM THE PISTON OF THE SYRINGE, AS SHOWN IN THE PICTURES.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1173790 SYRINGE NON STERILE 50ML LL SYRINGE FMF BECTON DICKINSON, S.A. 1904352

Patients

Seq Age Sex Outcome Treatment
1 Other