FDA Adverse Event Malfunction Summary report: N

BD DISCARDIT II 20 ML SYRINGE

MDR report key: 10705007 · Received October 20, 2020

Report

Report Number
3002682307-2020-00316
Event Type
Malfunction
Date Received
October 20, 2020
Date of Event
September 15, 2020
Report Date
December 30, 2020
Manufacturer
BECTON DICKINSON, S.A.
Product Code
FMF
PMA / PMN Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AU
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

THE FOLLOWING FIELDS WERE UPDATED DUE TO ADDITIONAL INFORMATION: D10: DEVICE AVAILABLE FOR EVAL: YES. D10: RETURNED TO MANUFACTURER ON: 2020-11-24. H6: INVESTIGATION SUMMARY: A DEVICE HISTORY RECORD REVIEW WAS PERFORMED FOR PROVIDED LOT NUMBER 2003251 AND THE REVIEW DID NOT REVEAL ANY DETECTED QUALITY ISSUES DURING THE PRODUCTION PROCESS THAT COULD HAVE CONTRIBUTED TO THIS REPORTED INCIDENT. TO AID IN THE INVESTIGATION OF THIS ISSUE, ONE PHYSICAL SAMPLE WAS RETURNED FOR EVALUATION BY OUR QUALITY ENGINEER TEAM. THROUGH EXAMINATION OF THE SAMPLE, LEAKAGE WAS OBSERVED THROUGH THE PLUNGER ROD. THROUGH MAGNIFIED INSPECTION, DAMAGE WAS OBSERVED TO THE PLUNGER ROD LIP COMPONENT. THIS TYPE OF DAMAGE CAN BE PRODUCED DURING THE HANDLING OF THE PRODUCT THROUGH THE MANUFACTURING PROCESS OR IN THE PLUNGER ASSEMBLY MACHINE. DAMAGE TO THE PLUNGER LIP COMPONENT CAN RESULT IN LEAKAGE.

Description of Event or Problem · 0

IT WAS REPORTED THAT BD DISCARDIT¿ II 20 ML SYRINGE LEAKED BEFORE USE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: RECENTLY, WE HAVE OFTEN NOTICED THE LEAKAGE OF DISCARDIT SYRINGES 20ML (WHEN ASPIRATING).

Additional Manufacturer Narrative · 1

(B)(6). A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED, BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT BD DISCARDIT¿ II 20 ML SYRINGE LEAKED BEFORE USE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: RECENTLY, WE HAVE OFTEN NOTICED THE LEAKAGE OF DISCARDIT SYRINGES 20ML (WHEN ASPIRATING).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1173697 BD DISCARDIT II 20 ML SYRINGE SYRINGE FMF BECTON DICKINSON, S.A. 2003251

Patients

Seq Age Sex Outcome Treatment
1 Other