BD RAPID DETECTION OF SARS-COV-2 VERITOR
Report
- Report Number
- 1119779-2020-00740
- Event Type
- Malfunction
- Date Received
- October 20, 2020
- Date of Event
- September 28, 2020
- Report Date
- October 14, 2021
- Manufacturer
- BECTON, DICKINSON & CO. (SPARKS)
- Product Code
- QKP
- PMA / PMN Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
H6: INVESTIGATION SUMMARY: THIS STATEMENT IS TO SUMMARIZE THE INVESTIGATION RESULTS REGARDING THE COMPLAINT THAT ALLEGES FALSE POSITIVE RESULTS WHEN USING KIT RAPID DETECTION OF SARS-COV-2 VERITOR (MATERIAL # 256082), BATCH NUMBER 0233466. BD QUALITY PERFORMS A SYSTEMATIC APPROACH TO INVESTIGATE FALSE POSITIVE/DISCREPANT RESULTS COMPLAINTS. THIS APPROACH INVOLVES REVIEW OF MANUFACTURING BATCH HISTORY RECORDS, TESTING OF RETENTION SAMPLES, AND TESTING OF CUSTOMER RETURNED SAMPLES, IF APPLICABLE. INITIATED AN INVESTIGATION AND TESTING WERE PERFORMED ON THE BATCH NUMBER PROVIDED. THE REPORTED ISSUE WAS UNABLE TO BE CONFIRMED. HOWEVER, THERE IS A TREND AGAINST FALSE POSITIVE RESULTS. BD HAS INITIATED CAPA (CORRECTIVE AND PREVENTIVE ACTION) # 1878253 TO FURTHER INVESTIGATE. THE ROOT CAUSE COULD NOT BE IDENTIFIED. BD QUALITY WILL CONTINUE TO CLOSELY MONITOR FOR TRENDS.
IT WAS REPORTED WHILE USING BD RAPID DETECTION OF SARS-COV-2 VERITOR ASSAY FALSE POSITIVE RESULTS WERE OBTAINED. REPEAT TESTS WERE PERFORMED USING A PCR TEST METHOD AND THE RESULTS WERE NEGATIVE. THE CUSTOMER STATED THE PATIENTS TESTED WERE ASYMPTOMATIC WITH NO KNOWN EXPOSURE. THIS TEST IS NOT INTENDED FOR USE ON ASYMPTOMATIC PATIENTS AND WAS THEREFORE USED OFF LABEL. PATIENTS WERE QUARANTINED AND EXPERIENCED EMOTIONAL DISTRESS. NO OTHER REPORT OF PATIENT IMPACT. EUA (B)(4).
(B)(4). A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED. (B)(4).
IT WAS REPORTED WHILE USING BD RAPID DETECTION OF SARS-COV-2 VERITOR ASSAY FALSE POSITIVE RESULTS WERE OBTAINED. REPEAT TESTS WERE PERFORMED USING A PCR TEST METHOD AND THE RESULTS WERE NEGATIVE. THE CUSTOMER STATED THE PATIENTS TESTED WERE ASYMPTOMATIC WITH NO KNOWN EXPOSURE. THIS TEST IS NOT INTENDED FOR USE ON ASYMPTOMATIC PATIENTS AND WAS THEREFORE USED OFF LABEL. PATIENTS WERE QUARANTINED AND EXPERIENCED EMOTIONAL DISTRESS. NO OTHER REPORT OF PATIENT IMPACT. (B)(4).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1173939 | BD RAPID DETECTION OF SARS-COV-2 VERITOR | CORONAVIRUS ANTIGEN DETECTION SYSTEM | QKP | BECTON, DICKINSON & CO. (SPARKS) | 0233466 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |