FDA Adverse Event
Malfunction
Summary report: N
GE OEC 9800
MDR report key: 1070297
·
Received July 9, 2008
Report
- Report Number
- 1720753-2008-20549
- Event Type
- Malfunction
- Date Received
- July 9, 2008
- Date of Event
- March 31, 2008
- Report Date
- April 10, 2008
- Manufacturer
- GE OEC MEDICAL SYSTEM INC.
- Product Code
- JAA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
A GE SERVICE REP PERFORMED AN ON SITE INVESTIGATION. THE COLLIMATOR WAS FOUND CLOSED. THE WIRING TO THE COLLIMATOR WAS REPAIRED DURING THE SERVICE CALL. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED, AND PUT BACK INTO SERVICE.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE 9800 SYSTEM HAD A LINE IN THE IMAGE. NO PT INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | GE OEC 9800 | FLUOROSCOPIC X-RAY | JAA | GE OEC MEDICAL SYSTEM INC. | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |