FDA Adverse Event
Malfunction
Summary report: N
GE OEC 9600
MDR report key: 1070291
·
Received July 9, 2008
Report
- Report Number
- 1720753-2008-20543
- Event Type
- Malfunction
- Date Received
- July 9, 2008
- Date of Event
- March 13, 2008
- Report Date
- April 10, 2008
- Manufacturer
- GE OEC MEDICAL SYSTEM INC.
- Product Code
- JAA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NJ, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
A GE SERVICE REP PERFORMED AN ON SITE INVESTIGATION. THE BOARDS WERE RE-SEATED. THE SOFTWARE WAS RE-INSTALLED AND THE INTERCONNECT CABLE WAS REPLACED DURING THE SERVICE CALL. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED, AND PUT BACK INTO SERVICE.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE 9600 SYSTEM WOULD INTERMITTENTLY NOT BOOT UP OR X-RAY. NO PT INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | GE OEC 9600 | FLUOROSCOPIC X-RAY | JAA | GE OEC MEDICAL SYSTEM INC. | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |