FDA Adverse Event Malfunction Summary report: N

9800

MDR report key: 1070281 · Received July 9, 2008

Report

Report Number
1720753-2008-20515
Event Type
Malfunction
Date Received
July 9, 2008
Date of Event
March 31, 2008
Report Date
April 8, 2008
Manufacturer
GE OEC MEDICAL SYSTEMS INC.
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KS, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THE SERVICE REP CONFIRMED ARCHING OF HIGH VOLTAGE CANDLES, CLEANED AND REGREASED CANDLES TESTED SYSTEM. SYSTEM OPERATES AS INTENDED.

Description of Event or Problem · 1

CUSTOMER STATES THAT SYSTEM WILL NOT FLUORO AND THEY WERE ALSO HEARING NOISE WHICH WAS POSSIBLY ARCHING. NO PT INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 9800 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS INC. NA NA

Patients

Seq Age Sex Outcome Treatment
1