FDA Adverse Event Malfunction Summary report: N

AFFINITY 4

MDR report key: 1070242 · Received June 30, 2008

Report

Report Number
1824206-2008-02748
Event Type
Malfunction
Date Received
June 30, 2008
Date of Event
February 27, 2008
Report Date
February 27, 2008
Manufacturer
HILL-ROM RITTER
Product Code
HDD
Removal / Correction Number
Z-0617-2008
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE HILL-ROM FIELD INVESTIGATION INDICATED THE BRAKES WOULD NOT HOLD ON THE BED. THE TECHNICIAN FOUND THE BRAKE DETENT INSTALLED UPSIDE DOWN. THE BRAKE DETENT WAS INSTALLED IN THE CORRECT POSITION TO REPAIR THE BED.

Description of Event or Problem · 1

BRAKES WILL NOT HOLD ON THE BED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 AFFINITY 4 BIRTHING BED HDD HILL-ROM RITTER 3700B

Patients

Seq Age Sex Outcome Treatment
1