FDA Adverse Event Malfunction Summary report: N

AFFINITY 4

MDR report key: 1070241 · Received June 30, 2008

Report

Report Number
1824206-2008-02749
Event Type
Malfunction
Date Received
June 30, 2008
Date of Event
November 5, 2007
Report Date
November 5, 2007
Manufacturer
HILL-ROM RITTER
Product Code
HDD
Removal / Correction Number
Z-0617-2008
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE HILL-ROM FIELD INVESTIGATION INDICATED THE BRAKES WOULD NOT HOLD ON THE BED. AFTER UNSUCCESSFUL ATTEMPTS TO ADJUST THE BRAKES IN THE FIELD, HILL-ROM ENGINEERING REQUESTED THE BED BE RETURNED. ENGINEERING FOUND THAT THE BRAKES, AS THEY WERE ADJUSTED IN THE FIELD, DID NOT HOLD PROPERLY WHEN THE BED WAS PUSHED AND TUGGED ON. ENGINEERING PROCEEDED TO READJUST THE CASTERS PER PROCEDURE AND WERE ABLE TO GET THE BRAKES TO HOLD PROPERLY. ENGINEERING CONCLUDED THE CASTERS WERE NOT ADJUSTED PROPERLY IN THE FIELD. MOD 373 IS A FIELD CORRECTIVE ACTION WHICH REPLACES THE BRAKE DETENT MECHANISM AND ADJUSTS ALL FOUR CASTERS OF THE AFFINITY 4 BIRTHING BED. THIS FAILURE OCCURRED FOLLOWING THE CORRECTION OF THIS UNIT.

Description of Event or Problem · 1

BRAKES WILL NOT STAY LOCKED ON THE BED.