FDA Adverse Event Malfunction Summary report: N

AFFINITY 4

MDR report key: 1070229 · Received June 30, 2008

Report

Report Number
1824206-2008-02622
Event Type
Malfunction
Date Received
June 30, 2008
Date of Event
December 28, 2007
Report Date
December 28, 2007
Manufacturer
HILL-ROM RITTER
Product Code
HDD
Removal / Correction Number
Z-0617-2008
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE HILL-ROM FIELD INVESTIGATION INDICATED TO BRAKE PEDAL WOULD NOT LOCK IN THE BRAKE POSITION. THE HILL-ROM FIELD TECHNICIAN ADJUSTED THE CASTERS TO REPAIR THE BED. MOD 373 IS A FIELD CORRECTIVE ACTION WHICH REPLACES THE BRAKE DETENT MECHANISM AND ADJUSTS ALL FOUR CASTERS OF THE AFFINITY 4 BIRTHING BED. THIS FAILURE OCCURRED FOLLOWING THE CORRECTION OF THIS UNIT.

Description of Event or Problem · 1

ALLEGED THE BRAKE WAS SET ON THIS BED, AND IT JUMPED INTO THE NEUTRAL POSITION WHILE THE PATIENT WAS ENTERING THE BED. THE PATIENT SLIPPED BUT WAS BEING ASSISTED, AND DID NOT FALL.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 AFFINITY 4 BIRTHING BED HDD HILL-ROM RITTER 3700B

Patients

Seq Age Sex Outcome Treatment
1 UNK