FDA Adverse Event Malfunction Summary report: N

AFFINITY 4

MDR report key: 1070217 · Received June 30, 2008

Report

Report Number
1824206-2008-02583
Event Type
Malfunction
Date Received
June 30, 2008
Date of Event
September 13, 2007
Report Date
September 13, 2007
Manufacturer
HILL-ROM RITTER
Product Code
HDD
Removal / Correction Number
Z-0617-2008
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE HILL-ROM FIELD INVESTIGATION INDICATED THE BRAKE PEDAL WOULD MIGRATE OUT OF THE BRAKE POSITION. THE HILL-ROM FIELD TECH ADJUSTED THE CASTERS TO REPAIR THE BED. MOD 373 IS A FIELD CORRECTIVE ACTION WHICH REPLACES THE BRAKE DETENT ASSEMBLY. THIS FAILURE OCCURRED FOLLOWING THE CORRECTION OF THIS UNIT.

Description of Event or Problem · 1

THE BRAKE WOULD "POP" OUT OF THE BRAKE POSITION IF THE BED WAS BUMPED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 AFFINITY 4 BIRTHING BED HDD HILL-ROM RITTER 3700B

Patients

Seq Age Sex Outcome Treatment
1