FREESTYLE FREEDOM
MDR report key: 1070206
·
Received July 3, 2008
Report
- Report Number
- 2954323-2008-02200
- Event Type
- Malfunction
- Date Received
- July 3, 2008
- Date of Event
- June 6, 2008
- Report Date
- July 3, 2008
- Manufacturer
- ABBOTT DIABETES CARE INC, USA
- Product Code
- LFR
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
THE CUSTOMER'S PRODUCT HAS BEEN REQUESTED BACK FOR AN INVESTIGATION. A FOLLOW-UP REPORT WILL BE FILED ONCE INVESTIGATION RESULTS ARE AVAILABLE. NOTE: IT SHOULD BE NOTED THAT THIS METER DOES NOT GIVE A VALUE FOR READINGS GREATER THAN 500 MG/DL. A "HI" DISPLAY MESSAGE INDICATES A READING GREATER THAN 500 MG/DL.
Description of Event or Problem · 1
CUSTOMER REPORTED RECEIVING ERRATIC READINGS ON THEIR FREESTYLE FREEDOM BLOOD GLUCOSE MONITOR WITHIN 10 MINUTES. RESULTS OF 501 MG/DL AND 107 MG/DL WERE PLOTTED ON A PARKES ERROR GRID. THE RESULTS FELL INTO THE "C" ZONE SHOWING THE DIFFERENCE IN VALUES TO BE CLINICALLY SIGNIFICANT. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY OR MISTREATMENT ASSOCIATED WITH THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | FREESTYLE FREEDOM | BLOOD GLUCOSE MONITORING SYSTEM | LFR | ABBOTT DIABETES CARE INC, USA | NA | 0801715 |