FDA Adverse Event Malfunction Summary report: N

INFINITI VISION SYSTEM

MDR report key: 1070131 · Received July 3, 2008

Report

Report Number
2028159-2008-00247
Event Type
Malfunction
Date Received
July 3, 2008
Date of Event
June 9, 2008
Report Date
June 9, 2008
Manufacturer
ALCON - IRIVINE TECHNOLOGY CENTER
Product Code
HQC
PMA / PMN Number
K021566
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE CUSTOMER DID NOT REQUEST SERVICE. THE INSTRUMENT SETTINGS PROVIDED WERE REVIEWED AND WERE DETERMINED THAT FROM THE SETTINGS ALONE, THEY COULD NOT HAVE CONTRIBUTED TO THE CORNEAL BURN. NO SAMPLES WERE RETURNED FOR EVALUATION, AND NO CONCLUSIONS COULD BE MADE FROM THE INFORMATION GIVEN. THEREFORE, THE ROOT CAUSE OF THE REPORTED EVENT IS UNKNOWN, AND COULD NOT BE DETERMINED. CORNEAL BURN IS AN ISSUE THAT IS OCCASIONALLY REPORTED WITH CATARACT SURGERY. ACCORDING TO THE ECRI HEALTH DEVICES, HAZARD UPDATE: SCLERAL AND CORNEAL BURNS DURING PHACOEMULSIFICATION, NOVEMBER 1996, VOL. 25, NO. 11: 426-431, MOST CORNEAL BURNS CAN BE TRACED TO ISSUES RELATED TO SURGICAL TECHNIQUE AND NOT TO MALFUNCTIONING EQUIPMENT. A LETTER AND COPY OF THIS INDUSTRY ARTICLE WAS PROVIDED TO THE CUSTOMER.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT TWO CORNEAL BURNS HAD OCCURED. THE SURGEON MADE A 2.6 MM INCISION, AND USED A 0.9 MM TAPERED TIP. THE SURGEON STATED THAT 2 OUT OF THE 5 CASES INVOLVED A CORNEAL BURN, AND THEY HAPPENED ON THE FIRST AND FOURTH CASES. THE BURNS WERE NOTED TO BE MILD. ONE STITCH WAS USED IN ALL 5 PATIENTS, AS THE DOCTOR TYPICALLY USES A STITCH. THE CUSTOMER STATED THE PATIENTS WERE EXPECTED TO DO WELL. THIS REPORT IS FOR ONE PATIENT; FOR ADDITIONAL PATIENT SEE MFG. REPORT #: 2028159-2008-00246.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 INFINITI VISION SYSTEM OPHTHALMIC SURGERY SYSTEM HQC ALCON - IRIVINE TECHNOLOGY CENTER INFINITI NA

Patients

Seq Age Sex Outcome Treatment
1 NI