MONOPOLAR CURVED SCISSORS INSTRUMENT
Report
- Report Number
- 2955842-2008-01111
- Event Type
- Malfunction
- Date Received
- July 3, 2008
- Report Date
- July 3, 2008
- Manufacturer
- INTUITIVE SURGICAL, INC.
- Product Code
- GEI
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- OTHER
Narratives
THE INSTRUMENT WAS RETURNED AND EVALUATED. PER THE CUSTOMER REPORTED COMPLAINT, ENGINEERING CONFIRMED THE DISTAL END OF THE MAIN TUBE HAS VARIOUS DEEP SCRATCHES CONCENTRATED ON ONE SIDE OF THE TUBE WITH MATERIAL REMOVED. THE SCRATCHES ARE NOT ALIGNED WITH THE TUBE AXIS. BASED ON THE LOCATION AND APPEARANCE, THE DAMAGE WAS MOST LIKELY CAUSED BY INSTRUMENT COLLISIONS. NO OTHER DAMAGE FOUND. THE ENDOWRIST INSTRUMENT INSTRUCTIONS FOR USE (IFU) SPECIFICALLY STATES: GENERAL PRECAUTIONS AND WARNINGS: HANDLE INSTRUMENTS WITH CARE. AVOID MECHANICAL SHOCK OR STRESS THAT CAN CAUSE DAMAGE TO THE INSTRUMENTS. DO NOT USE AN INSTRUMENT TO CLEAN DEBRIS FROM ANOTHER INSTRUMENT INTRAOPERATIVELY. THIS MAY RESULT IN DAMAGE TO THE INSTRUMENTS OR OTHER UNINTENDED CONSEQUENCES, SUCH AS DISCONNECTION OF THE INSTRUMENT TIP.
IT WAS REPORTED THAT THE MONOPOLAR CURVED SCISSORS INSTRUMENT HAS "METAL SLIVERS" COMING OFF AFTER STERILIZATION PROCESS. NO ADDITIONAL INFORMATION PROVIDED. NO PATIENT HARM, ADVERSE OUTCOME OR INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | MONOPOLAR CURVED SCISSORS INSTRUMENT | ELECTROSURGICAL INSTRUMENT | GEI | INTUITIVE SURGICAL, INC. | 420179-08 | 0712281 004 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | ACCESSORIES| ELECTROSURGICAL UNIT| DA VINCI S SURGICAL SYSTEM |