FDA Adverse Event Malfunction Summary report: N

8MM MCS TIP COVER ACCESSORY

MDR report key: 1070036 · Received July 3, 2008

Report

Report Number
2955842-2008-01125
Event Type
Malfunction
Date Received
July 3, 2008
Report Date
July 3, 2008
Manufacturer
INTUITIVE SURGICAL, INC.
Product Code
GEI
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE TIP COVER ACCESSORY WAS RETURNED AND EVALUATED. PER THE CUSTOMER REPORTED COMPLAINT, ENGINEERING FOUND MULTIPLE HOLES IN THE SILICONE WITH LOCALIZED MELTING AROUND ALL HOLES THAT IS INDICATIVE OR ARCING. ALL OF THE DAMAGE IS CONTAINED WITHIN ONE HALF OF THE SILICONE, BETWEEN THE TWO PARTING LINES. THE HOLES ARE ALL UNDER .030 INCHES WITH VARYING SHAPES, ROUND AND OBLONG, AND ARE LOCATED FROM .275 INCHES TO .530 INCHES ABOVE THE SILICONE TO SLEEVE INTERFACE. MULTIPLE HOLES INDICATE MULTIPLE ARCING EVENTS OCCURRED. THE TIP COVER ALSO HAS VARIOUS MECHANICAL TEARS IN THE GENERAL VICINITY OF THE HOLES, LIKELY DUE TO INSTRUMENT COLLISIONS OR REPEATED INSTRUMENT WRIST ARTICULATIONS. ENGINEERING CONCLUDED THAT ARCING IS LIKELY DUE TO INSUFFICIENT SILICONE DIELECTRIC STRENGTH, WITH STRENGTH WEAKENED DUE TO INSTRUMENT COLLISIONS. HIGH GENERATOR SETTINGS MAY HAVE ALSO CONTRIBUTED TO ARCING. THE ENDOWRIST INSTRUMENTS INSTRUCTIONS FOR USE (IFU) SPECIFICALLY STATES: GENERAL PRECAUTIONS AND WARNINGS: HANDLE INSTRUMENTS WITH CARE. AVOID MECHANICAL SHOCK OR STRESS THAT CAN CAUSE DAMAGE TO THE INSTRUMENTS. DO NOT USE AN INSTRUMENT TO CLEAN DEBRIS FROM ANOTHER INSTRUMENT INTRAOPERATIVELY. THIS MAY RESULT IN DAMAGE TO THE INSTRUMENTS OR OTHER UNINTENDED CONSEQUENCES, SUCH AS DISCONNECTION OF THE INSTRUMENT TIP.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THE TIP COVER ACCESSORY HAS HOLES ON THE SIDE. NO PATIENT HARM, ADVERSE OUTCOME OR INJURY WAS REPORTED. THE FOLLOWING MEDWATCH REPORT WAS SENT TO THE FDA RELATING TO THIS EVENT: 2955842-2008-01126.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 8MM MCS TIP COVER ACCESSORY ELECTROSURGICAL INSTRUMENT ACCESSORY GEI INTUITIVE SURGICAL, INC. 400180-05 C7355

Patients

Seq Age Sex Outcome Treatment
1 ELECTROSURGICAL UNIT| DA VINCI S SURGICAL SYSTEM| MONOPOLAR CURVED SCISSORS| INSTRUMENT