FDA Adverse Event Malfunction Summary report: N

PROGRAMMING WAND

MDR report key: 1070003 · Received July 2, 2008

Report

Report Number
1644487-2008-01502
Event Type
Malfunction
Date Received
July 2, 2008
Date of Event
June 3, 2008
Report Date
June 3, 2008
Manufacturer
CYBERONICS, INC.
Product Code
LYJ
PMA / PMN Number
P970003
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

DEVICE FAILURE OCCURRED, BUT DID NOT CAUSE OR CONTRIBUTE TO DEATH OR SERIOUS INJURY.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE SITE WAS EXPERIENCING PROBLEMS WITH THE PROGRAMMING SYSTEM. THE WAND COULD NOT BE RULED OUT AS A CONTRIBUTORY FACTOR DURING TROUBLESHOOTING. THE WAND WAS SUBSEQUENTLY RETURNED TO MFR FOR ANALYSIS. ANALYSIS OF THE DEVICE REVEALED THAT THE SERIAL DATA CABLE, WHICH PRODUCED COMMUNICATION ERRORS, HAD INTERMITTENT CONDUCTORS. A KNOWN GOOD BENCH SERIAL DATA CABLE WAS SUBSTITUTED AND ALL COMMUNICATIONS ERRORS CLEARED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PROGRAMMING WAND LYJ CYBERONICS, INC. 201

Patients

Seq Age Sex Outcome Treatment
1