FDA Adverse Event Malfunction Summary report: N

KINGFISHER DNA EXTRACTOR THERMO FISHER

MDR report key: 10698382 · Received October 16, 2020

Report

Report Number
MW5097305
Event Type
Malfunction
Date Received
October 16, 2020
Report Date
March 3, 2025
Manufacturer
THERMO FISHER SCIENTIFIC (ASHEVILLE) LLC
Product Code
LXG
Product Problem
Yes
Report Source
Voluntary report
Reporter Location
CT, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Description of Event or Problem · 0

ADDITIONAL INFORMATION RECEIVED FOR REPORT MW5097305 TO ADD AN ADDITIONAL DEVICE, UPDATE SERIAL NUMBER AND TO CHANGE PROCODE. REFERENCE REPORT: MW5167161. PROCODE: LXG.

Description of Event or Problem · 1

THE KINGFISHER SPLANKER WHEN EXTRACTING DNA. THIS RESULTS IN CONTAMINATION OF THE OTHER SAMPLES IN THE PLATE WITH POSITIVE CONTROLS THAT ARE BEING EXTRACTED. THIS RESULTS IN FALSE POSITIVE REPORTING FOR COVID-19 RESULT. THERE ARE ALSO TIP CARE ISSUES WITH THE MACHINE SPECIALLY WITH THE ALTERNATE PLANTATION THAT THE MANUFACTURER IS SELLING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1152749 KINGFISHER DNA EXTRACTOR THERMO FISHER EQUIPMENT, LABORATORY, GENERAL PURPOSE, LABELED OR PROMOTED FOR A SPECIFIC MEDIC LXG THERMO FISHER SCIENTIFIC (ASHEVILLE) LLC KINGFISHERFLEX
1152750 KINGFISHER DNA EXTRACTOR THERMO FISHER ANALYZER, CHROMOSOME, AUTOMATED LXG THERMO FISHER

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown