STEM: QUADRA-H CEMENTLESS, HA COATED STD STEM SIZE 0, SHORT NECK

MDR report key: 10697594 · Received October 19, 2020

Report

Report Number
3005180920-2020-00724
Event Type
Injury
Date Received
October 19, 2020
Date of Event
September 28, 2020
Report Date
October 19, 2020
Manufacturer
MEDACTA INTERNATIONAL SA
Product Code
JDI
UDI-DI
07630030802607
PMA / PMN Number
K082792
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
BE
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 1

BATCH REVIEW PERFORMED ON (B)(6) 2020. LOT 1905425: 50 ITEMS MANUFACTURED AND RELEASED ON 08-OCT-2019. EXPIRATION DATE: 2024-09-24. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, 21 ITEMS OF THE SAME LOT HAVE BEEN ALREADY SOLD WITHOUT ANY OTHER SIMILAR REPORTED EVENT.

Description of Event or Problem · 1

PRIMARY SURGERY PERFORMED SUCCESSFULLY IN THE MORNING. IN THE AFTERNOON THE SURGEON NOTICED THAT THE PATIENT WAS IN PAIN AND LEGG WAS SHORTER THAN THE OTHER SUDDENLY. HE TOOK A CT WHERE THE STEM WAS SUBSIDED AND IT SHOWED A PERI PROSTHETIC FRACTURE, SUBSIDENCE HAPPENED AFTER CAUSING THE FRACTURE. PATIENT HAS A VERY OSTEOPOROTIC BONE QUALITY. REVISION SURGERY PERFORMED 7 DAYS AFTER PRIMARY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1164974 STEM: QUADRA-H CEMENTLESS, HA COATED STD STEM SIZE 0, SHORT NECK HIP CEMENTLESS STEM JDI MEDACTA INTERNATIONAL SA 01.12.20SN 1905425 07630030802607

Patients

Seq Age Sex Outcome Treatment
1 73 YR Required Intervention