FDA Adverse Event Injury Summary report: N

CRYOSPRAY ABLATION SYSTEM

MDR report key: 1069515 · Received May 7, 2008

Report

Report Number
127116-2008-00005
Event Type
Injury
Date Received
May 7, 2008
Date of Event
April 7, 2008
Report Date
May 7, 2008
Manufacturer
CSA MEDICAL, INC.
Product Code
GEH
PMA / PMN Number
k060555
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

PRIOR TO THERAPY, THE PHYSICIAN PLACED AN INK TATTOO IN THE ESOPHAGEAL WALL USING A SCLERA THERAPY NEEDLE NEAR THE TREATMENT SITE. IT IS THE OPINION OF THE TREATING PHYSICIAN THAT NITROGEN GAS INFILTRATED THE ESOPHAGEAL WALL THROUGH THE NEEDLE PUNCTURE SITE INTRODUCED DURING THE TATTOO PROCESS. INVESTIGATION BY CSA MEDICAL CONCLUDED THAT THIS WAS THE MOST LIKELY CAUSE OF THE MICRO-PERFORATION. TATTOOING IS A PROCEDURE USED TO MARK TREATMENT OR DISEASE SITES, OFTEN DURING CLINICAL RESEARCH. THE PROCESS CAN COMPROMISE THE INTEGRITY OF THE GASTROINTESTINAL TRACT AND SHOULD NOT BE DONE PRIOR TO THERAPY. THIS WAS CLARIFIED WITH THE TREATING PHYSICIAN.

Description of Event or Problem · 1

A PT DEVELOPED AN ESOPHAGEAL MICROPERFORATION WITH PNEUMOMEDIASTINUM AND SUBCUTANEOUS EMPHYESMA (AIR) IMMEDIATELY AFTER TREATMENT WITH THE CSA CRYOSPRAY. A GASTROGRAFFIN SWALLOW WAS NEGATIVE FOR A PERFORATION (I.E. SHOWED NO LEAK.) THE PT WAS HOSPITALIZED AND MANAGED WITH IV ANTIBIOTICS. THE CONDITION RESOLVED COMPLETELY AND THE PT WAS DISCHARGED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CRYOSPRAY ABLATION SYSTEM CRYOSPRAY GEH CSA MEDICAL, INC. CC2-NAM *

Patients

Seq Age Sex Outcome Treatment
1 * Hospitalization