FDA Adverse Event Malfunction Summary report: N

NARROW RIGHT ANGLE LARGE HEX DRIVER TIP 30MM

MDR report key: 10693779 · Received October 16, 2020

Report

Report Number
0001038806-2020-01590
Event Type
Malfunction
Date Received
October 16, 2020
Date of Event
August 30, 2020
Report Date
January 15, 2021
Manufacturer
BIOMET 3I
Product Code
NDP
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

ONE NARROW RIGHT ANGLE LARGE HEX DRIVER TIP 30MM (RASH8N) WAS RETURNED FOR INVESTIGATION. VISUAL INSPECTION OF THE AS RETURNED PRODUCT IDENTIFIED THAT THE DRIVER TIP AND LATCHLOCK ARE MODERATELY WORN, WHILE THE BODY OF THE DRIVER IS FRACTURED LATERALY. THE REPORTED EVENT COULD NOT BE RECREATED DUE TO THE NATURE OF THE DENTAL DEVICE AND EVENT (FRACTURE). DHR REVIEW WAS COMPLETED FOR THE SUBJECT LOT NUMBER 1152175. IT WAS CONFIRMED THAT ALL OPERATIONS AND INSPECTIONS WERE EXECUTED AS PER APPLICABLE PROCEDURE. NO DEVIATIONS OR NON-CONFORMANCES, WHICH COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED EVENT WAS NOTED AS PART OF THE DHR. LOT WAS INSPECTED AND PASSED ALL ACCEPTANCE CRITERIA BY QA. COMPLAINT HISTORY REVIEW WAS PERFORMED FOR THE REPORTED LOT NUMBER (1152175) FOR SIMILAR EVENT AND NO OTHER COMPLAINT WAS IDENTIFIED. NOVEMBER POST MARKET TRENDING WAS REVIEWED AND THERE WERE NO ACTIONABLE EVENTS OR CORRECTIVE ACTIONS FOR THE REPORTED EVENT (COMPONENT FRACTURE) OR PRODUCT (RASH8N). THEREFORE, BASED ON THE AVAILABLE INFORMATION, DEVICE MALFUNCTION HAS OCCURRED AND THE REPORTED EVENT WAS CONFIRMED.

Additional Manufacturer Narrative · 0

THIS REPORT HAS BEEN RELAYED TO CORRECT AND ERROR IN THE PREVIOUSLY REPORTED SUBMISSION (MFR# 0001038806-2020-01590-1).

Description of Event or Problem · 0

NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.

Description of Event or Problem · 0

NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT. SEE H10 FOR CORRECTION DETAILS.

Additional Manufacturer Narrative · 1

ZIMMER BIOMET (B)(4). AGE: NOT PROVIDED. PATIENT WEIGHT: NOT PROVIDED. DEVICE NOT RETURNED.

Description of Event or Problem · 1

IT WAS REPORTED THAT DRIVER TIP (RASH8N) FRACTURED DURING A SECOND NORMAL USED. DOCTOR DID NOT FIND ANY PRODUCT QUALITY AND NO HEALTH HAZARD WAS REPORTED.