FDA Adverse Event Malfunction Summary report: N

FINESSE

MDR report key: 10693254 · Received October 16, 2020

Report

Report Number
3009974348-2020-00353
Event Type
Malfunction
Date Received
October 16, 2020
Date of Event
September 17, 2020
Report Date
September 18, 2020
Manufacturer
AMEDA, INC.
Product Code
HGX
UDI-DI
00810725023563
PMA / PMN Number
K973501
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

RETURNED AC ADAPTER, P/N 4100103, MODEL DYS612-090130W-1 LOT 2819 HAS VISIBLE DAMAGE TO PRONGS (BENT UPWARD) AND ABOVE PRONG LOCATION 120V HOUSING, WHICH IS MISSING A SIGNIFICANT SECTION. FAINT ODOR OF AN ELECTRONIC COMPONENT FAILURE/THERMAL EVENT NOTICED. RETURNED AC ADAPTER, P/N 4100103, MODEL DYS612-090130W-1 LOT 2819 HOUSING WAS THEN OPENED FOR INTERNAL INVESTIGATION. REMOVED AC ADAPTER PCB TO INSPECT BOTH SIDES OF PCB WHERE DAMAGED PRONG IS CONNECTED/LOCATED. MINOR CHARRING IS EVIDENT AROUND DAMAGED PRONG/PCB CONNECTION LOCATION WHERE THERMAL EVENT OCCURRED. ENGINEERING CONCLUSION EVIDENCE DOES NOT SUPPORT AN EXPLOSIVE EVENT. DAMAGE TO THE PRONGS AND THEIR CONNECTION TO THE PCB INDICATES THAT THE AC ADAPTER MOST LIKELY SUSTAINED DAMAGE PRIOR TO THE "SPARK" EVENT. UNABLE TO DETERMINE THE FAULT OF AC ADAPTER FAILURE/DAMAGE. TESTING OF RETURNED FINESSE PUMP WAS WAIVED DUE TO COMMUNICATION LOG INDICATING THE PUMP STILL FUNCTIONS. PRELIMINARY INSPECTION OF RETURNED FINESSE P/N #4100018, S/N# (B)(4) /SW MEETS SPECIFICATIONS.

Description of Event or Problem · 1

CUSTOMER CONTACTED AMEDA, INC. ON (B)(6) 2020 TO REPORT AN EVENT THAT OCCURRED RECENTLY WHEN USING THE FINESSE PUMP WITH THE AMEDA AC ADAPTER. CUSTOMER HAS BEEN PUMPING FOR 5 MONTHS UP TO 4 TIMES/DAY. SHE INITIALLY REPORTS HEARING A "TICKING" SOUND COMING FROM THE AC ADAPTER. UPON SUBSEQUENT USE, SHE SAID THE AC ADAPTER FELT WARM TO THE TOUCH. FOLLOWING A FEW MORE USES, SHE SAID THE AC ADAPTER "SPARKED AND SEEMED TO FALL APART". SHE SAYS THE PUMP WILL STILL WORK, BUT SHE IS VERY CONCERNED. CUSTOMER WAS INSTRUCTED NOT TO USE THIS PUMP ANY LONGER. A REPLACEMENT FINESSE AND AC ADAPTER WAS SHIPPED OVERNIGHT TO THE CUSTOMER.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1159033 FINESSE ELECTRIC BREAST PUMP HGX AMEDA, INC. 24502082 2819 00810725023563

Patients

Seq Age Sex Outcome Treatment
1 24 YR