FDA Adverse Event Malfunction Summary report: N

PRISMAFLO IIS

MDR report key: 10690473 · Received October 16, 2020

Report

Report Number
9617473-2020-00001
Event Type
Malfunction
Date Received
October 16, 2020
Manufacturer
STIHLER ELECTRONIC GMBH
Product Code
KOC
PMA / PMN Number
K082758
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THIS INCIDENT WAS CAUSED BY USER FAULT. PICTURES (PROVIDED BY THE HEALTH CARE FACILITY), THAT THERE HAS BEEN A MISUSE OF THE DEVICE. IN THE IFU AT FOUR DIFFERENT CHAPTERS THERE ARE PLACED WARNINGS - RISK OF OVERHEATING / DO NOT COVER (NOT EVEN PARTIALLY) THE HEATING PROFILE WHILE IN USE. ADDITIONALLY ON THE BLACK SHORT HOUSING OF THE HEATING PROFILE THERE IS PLACED A RED WARNING LABEL WITH THE SYMBOL "PROHIBITION: DO NOT COVER HEATING PROFILE - RISK OF OVERHEATING".

Description of Event or Problem · 1

IT WAS REPORTED THAT PHYSICIAN AND BEDSIDE NURSE ENTERED PATIENT'S ROOM TO DELIVER PATIENT CARE. WHEN THEY MOVED THE PILLOW ON THE BED BY THE PATIENT'S LEG, THEY OBSERVED THE BURNT AREA ON THE BED AND PILLOW. THEY POURED WATER ON THE BED TO EXTINGUISH IT. A WARMER LINE OF THE HEATER (PRISMAFLO II DEVICE) ATTACHED TO THE CVVH PRISMA FLEX MACHINE, THAT REWARMS BLOOD ON THE RETURN TO THE PATIENT, CAUSED A SMALL FIRE IN THE MATTRESS AND PILLOW. THE TUBING LINE WAS IN THE PATIENT BED AND WAS FOUND TO BE IN BETWEEN THE PILLOW AND THE MATTRESS AND BEGUN TO BURN.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1159149 PRISMAFLO IIS BLOOD WARMER KOC STIHLER ELECTRONIC GMBH PF2-WP33

Patients

Seq Age Sex Outcome Treatment
1