CARTO VIZIGO 8.5F BI-DIRECTIONAL GUIDING SHEATH - MEDIUM
Report
- Report Number
- 2029046-2020-01482
- Event Type
- Malfunction
- Date Received
- October 15, 2020
- Date of Event
- August 19, 2020
- Report Date
- September 17, 2020
- Manufacturer
- BIOSENSE WEBSTER INC
- Product Code
- DYB
- UDI-DI
- 10846835016277
- PMA / PMN Number
- K170997
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- OTHER
Narratives
THE BWI PRODUCT ANALYSIS LAB RECEIVED THE DEVICE FOR EVALUATION. THE PRODUCT INVESTIGATION WAS SUBSEQUENTLY COMPLETED. THE DEVICE WAS VISUALLY INSPECTED, AND THE HEMOSTATIC VALVE WAS FOUND DISLODGED INSIDE OF HUB. DILATOR WAS NOT RETURNED. IT WAS DETERMINED THAT THE ISSUE OBSERVED COULD BE RELATED TO THE INCORRECT INSERTION OF THE DILATOR INTO THE SHEATH CAUSING THE DISLODGMENT OF THE VALVE, SINCE A MICROSCOPE TESTING WAS PERFORMED AND EVIDENCE OF MECHANICAL DAMAGE WAS OBSERVED ON THE HEMOSTATIC VALVE. THIS DAMAGE ALSO SUGGEST THAT THE DILATOR WAS TRIED TO BE INTRODUCED NOT ON A STRAIGHT POSITION AS INDICATED ON THE ODP. ACCORDING TO THE ODP (OPTIMAL PERFORMANCE GUIDE), THERE ARE SOME PRECAUTIONS ON INSERTING THE DILATOR INTO THE VIZIGO SHEATH. ALWAYS INSERT A DILATOR STRAIGHT INTO THE CENTER OF THE SHEATH¿S VALVE TO PREVENT DAMAGE TO THE VALVE. DO NOT INSERT A DILATOR AT AN ANGLE, AS DAMAGE TO THE SHEATH VALVE MAY OCCUR. A DEVICE HISTORY RECORD EVALUATION WAS PERFORMED FOR THE FINISHED DEVICE (B)(4) NUMBER, AND NO INTERNAL ACTION RELATED TO THE COMPLAINT WAS FOUND DURING THE REVIEW. THE ISSUE REPORTED BY THE CUSTOMER WAS CONFIRMED. AND IT APPEARS TO BE RELATED TO THE INCORRECT INTRODUCTION OF THE VESSEL DILATOR. THE ODP (OPTIMAL DEVICE PERFORMANCE GUIDE) PROVIDE ADDITIONAL INSTRUCTIONS ON HOW TO INSERT THE DILATOR INTO THE SHEATH. IN ADDITION, THERE IS EVIDENCE THAT THE DEVICE WAS MANUFACTURED IN ACCORDANCE WITH DOCUMENTED SPECIFICATION AND PROCEDURES. IF ADDITIONAL INFORMATION IS RECEIVED REGARDING THIS EVENT, A SUPPLEMENTAL 3500A REPORT WILL BE SUBMITTED TO THE FDA. MANUFACTURER'S REFERENCE NUMBER: (B)(4).
A PATIENT UNDERWENT AN ATRIAL FIBRILLATION (AFIB) ABLATION PROCEDURE WITH A CARTO VIZIGO¿ 8.5F BI-DIRECTIONAL GUIDING SHEATH - MEDIUM FOR WHICH BIOSENSE WEBSTER¿S PRODUCT ANALYSIS LAB IDENTIFIED THAT THE HEMOSTATIC VALVE WAS DISLODGED INSIDE OF THE HUB. DURING THE PROCEDURE, THE DILATOR WAS NOT PROPERLY ADVANCING THROUGH THE HANDLE OF THE CARTO VIZIGO¿ 8.5F BI-DIRECTIONAL GUIDING SHEATH - MEDIUM. THE DILATOR COULD ONLY BE ADVANCED APPROXIMATELY 6 INCHES BEFORE IT COULD NOT ADVANCE ANY FURTHER. IT WAS REPORTED THAT THE DILATOR FELT AS IF IT "HIT A WALL". THE WIRES WERE ALSO OBSTRUCTED FROM ADVANCEMENT THROUGH THE SHEATH. THE SHEATH WAS REPLACED AND THE ISSUES WERE RESOLVED. THE CARTO 3 SYSTEM IS OPERATING PER SPECS AND IS NOT RESPONSIBLE FOR THE PRODUCT ISSUE. THE PROCEDURE CONTINUED, AND NO PATIENT CONSEQUENCES WERE REPORTED. THE HEMOSTATIC VALVE DISLODGMENT IS AN MDR-REPORTABLE ISSUE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1146720 | CARTO VIZIGO 8.5F BI-DIRECTIONAL GUIDING SHEATH - MEDIUM | INTRODUCER, CATHETER | DYB | BIOSENSE WEBSTER INC | D138502 | 00001302 | 10846835016277 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |