FDA Adverse Event Malfunction Summary report: N

ATELLICA SAMPLE HANDLER PRIME

MDR report key: 10680629 · Received October 14, 2020

Report

Report Number
2432235-2020-00414
Event Type
Malfunction
Date Received
October 14, 2020
Date of Event
September 22, 2020
Report Date
November 17, 2020
Manufacturer
SIEMENS HEALTHCARE DIAGNOSTICS INC.
Product Code
JJE
UDI-DI
00630414002439
PMA / PMN Number
K161954
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SZ
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

SIEMENS FILED THE INITIAL MDR 2432235-2020-00414 ON 14-OCT-2020. ADDITIONAL INFORMATION (23-OCT-2020): SIEMENS REVIEWED THE INFORMATION PROVIDED AND SYSTEM FILES. IT WAS DETERMINED THAT A JAM IN THE CUVETTELOADER OCCURRED, AND THE TEST SCHEDULED FOR SAMPLE 350445 WAS CANCELED BECAUSE CUVETTES WERE NOT AVAILABLE FOR PROCESSING. SIEMENS DETERMINED THAT THE ATELLICA SAMPLE HANDLER PRIME PERFORMED AS SPECIFIED AND IDENTIFIED NO PRODUCT ISSUE. THE CUSTOMER HAS RESTARTED THE PROCESS CONTROL COMPUTER (PCC) AND RESOLVED THE ISSUE. THE SYSTEM IS PERFORMING WITHIN SPECIFICATIONS. NO FURTHER EVALUATION OF THE DEVICE WAS REQUIRED.

Additional Manufacturer Narrative · 1

THE CUSTOMER CONTACTED THE SIEMENS CUSTOMER CARE CENTER (CCC) AND INFORMED THAT THE DELAY WAS 1-2 HOURS. SIEMENS IS INVESTIGATING THE EVENT.

Description of Event or Problem · 1

THE CUSTOMER NOTED THAT PUCKS ON THE ATELLICA SAMPLE HANDLER PRIME STOPPED MOVING WITHOUT ERROR, WHILE THE ATELLICA MODULE WAS IN THE READY OR THE PROCESSING STATE. ON THE FOLLOWING MORNING ((B)(6) 2020), BETWEEN 7:30 A.M. AND 8:00 A.M., THE CUSTOMER RESTARTED THE ATELLICA AND NOTED IT WAS IN THE READY STATE. THE CUSTOMER LOADED A STAT SAMPLE ON A PUCK ON THE ATELLICA SAMPLE HANDLER PRIME, AND THE PUCK STOPPED AT THE DEFAULT WAITING QUEUE (DWQ). THE CUSTOMER NOTED THAT THE STAT SAMPLE DID NOT PROCESS AND THAT OTHER SAMPLES LOADED THROUGH THE SAMPLE HANDLER CONNECT (SHC) WERE NOT PROCESSED AND REMAINED ON THE DWQ. THE CUSTOMER PERFORMED TROUBLESHOOTING, RESTARTED THE PROCESS CONTROL COMPUTER (PCC), AND RESOLVED THE ISSUE. THERE ARE NO REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE DELAY IN STAT TESTING AND REPORTING RESULTS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1145381 ATELLICA SAMPLE HANDLER PRIME ATELLICA SAMPLE HANDLER PRIME JJE SIEMENS HEALTHCARE DIAGNOSTICS INC. ATELLICA SAMPLE HANDLER PRIME 00630414002439

Patients

Seq Age Sex Outcome Treatment
1