FDA Adverse Event Malfunction Summary report: N

STERILE CLEANWIPE FOAM SWAB

MDR report key: 10676738 · Received October 13, 2020

Report

Report Number
MW5097197
Event Type
Malfunction
Date Received
October 13, 2020
Date of Event
October 2, 2020
Report Date
October 9, 2020
Manufacturer
FOAMTEC INTERNATIONAL CO., LTD.
Product Code
KXG
Product Problem
Yes
Report Source
Voluntary report
Reporter Location
LA, US
Reporter Occupation
PHYSICIAN

Narratives

Description of Event or Problem · 1

FOAM COATING AT THE TIP OF SWAB USED FOR COVID-19 TESTING CAME OFF OF THE SWAB SHAFT WHILE IN PATIENT'S NOSE. PATIENT USED NASAL SALINE LAVAGE TO SUCCESSFULLY REMOVE THE FOAM TIP. SEVERAL SWABS FROM THE SAME MANUFACTURER AND LOT NUMBER WERE DEFECTIVE, WITH FOAM TIP COMING OFF PRIOR TO USE IN PATIENTS. WE TESTED 100 SWABS FROM THE SAME LOT NUMBER TO SEE IF THE TIPS WERE EASILY REMOVED WITH LIGHT PRESSURE. NINE OF 100 WERE DEFECTIVE BY THIS TECHNIQUE. SWABS ARE FROM FOAMTEC INTERNATIONAL, PRODUCT #MP1301AST, LOT #0640202040. FDA SAFETY REPORT ID# (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1134279 STERILE CLEANWIPE FOAM SWAB APPLICATOR, ABSORBENT TIPPED, STERILE KXG FOAMTEC INTERNATIONAL CO., LTD. 0640202040
1134282 STERILE CLEANWIPE FOAM SWAB APPLICATOR, ABSORBENT TIPPED, STERILE KXG FOAMTEC INTERNATIONAL CO., LTD. 0640202040
1134283 STERILE CLEANWIPE FOAM SWAB APPLICATOR, ABSORBENT TIPPED, STERILE KXG FOAMTEC INTERNATIONAL CO., LTD. 0640202040
1134284 STERILE CLEANWIPE FOAM SWAB APPLICATOR, ABSORBENT TIPPED, STERILE KXG FOAMTEC INTERNATIONAL CO., LTD. 0640202040
1134285 STERILE CLEANWIPE FOAM SWAB APPLICATOR, ABSORBENT TIPPED, STERILE KXG FOAMTEC INTERNATIONAL CO., LTD. 0640202040
1134286 STERILE CLEANWIPE FOAM SWAB APPLICATOR, ABSORBENT TIPPED, STERILE KXG FOAMTEC INTERNATIONAL CO., LTD. 0640202040
1134287 STERILE CLEANWIPE FOAM SWAB APPLICATOR, ABSORBENT TIPPED, STERILE KXG FOAMTEC INTERNATIONAL CO., LTD. 0640202040
1134288 STERILE CLEANWIPE FOAM SWAB APPLICATOR, ABSORBENT TIPPED, STERILE KXG FOAMTEC INTERNATIONAL CO., LTD. 0640202040
1134289 STERILE CLEANWIPE FOAM SWAB APPLICATOR, ABSORBENT TIPPED, STERILE KXG FOAMTEC INTERNATIONAL CO., LTD. 0640202040

Patients

Seq Age Sex Outcome Treatment
1 Other