FDA Adverse Event Injury Summary report: N

HEATER-COOLER SYSTEM 3T

MDR report key: 10676166 · Received October 14, 2020

Report

Report Number
9611109-2020-00568
Event Type
Injury
Date Received
October 14, 2020
Date of Event
January 5, 2007
Report Date
October 12, 2020
Manufacturer
LIVANOVA DEUTSCHLAND
Product Code
DWC
Removal / Correction Number
Z-2076/2081-2015
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
IT
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THERE WAS NO KNOWN PATIENT INVOLVEMENT. SERIAL NUMBER IS UNKNOWN. THIS INFORMATION WILL BE PROVIDED IN A SUPPLEMENTAL REPORT IF MADE AVAILABLE. AS THE SERIAL NUMBER IS UNKNOWN, THE DEVICE MANUFACTURE DATE COULD NOT BE DETERMINED. THIS INFORMATION WILL BE PROVIDED IN A SUPPLEMENTAL REPORT IF MADE AVAILABLE. THE HEATER-COOLER 16-02-80 IS NOT DISTRIBUTED IN THE USA, AND IT IS SIMILAR TO HEATER-COOLER 16-02-85, WHICH IS DISTRIBUTED IN THE USA (510(K) NUMBER: K191402). LIVANOVA (B)(4) IMPLEMENTED A FIELD SAFETY NOTICE FOR DISINFECTION AND CLEANING OF HEATER-COOLER DEVICES. THE Z NUMBER IS Z-2076/2081-2015. LIVANOVA (B)(4) MANUFACTURES THE HEATER-COOLER SYSTEM 3T . THE INCIDENT OCCURRED IN (B)(6). LIVANOVA INITIATED AN INVESTIGATION. IF ANY ADDITIONAL INFORMATION PERTINENT TO THE REPORTED EVENT IS RECEIVED, IT WILL BE PROVIDED IN A SUPPLEMENTAL REPORT.

Description of Event or Problem · 1

THROUGH LITERATURE REVIEW LIVANOVA BECAME AWARE OF A (B)(6) YEARS OLD PATIENT INFECTED WITH MYCOBACTERIUM CHIMAERA. THE PATIENT WAS FOUND POSITIVE AFTER 12 YEARS AFTER A CARDIAC SURGERY IN 2007. THE IMPLICATION OF HEATER-COOLER SYSTEM 3T IS SUSPECTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1141570 HEATER-COOLER SYSTEM 3T CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS DWC LIVANOVA DEUTSCHLAND 16-02-80

Patients

Seq Age Sex Outcome Treatment
1 Other