CATALYS SYSTEM
Report
- Report Number
- 3005675890-2020-00052
- Event Type
- Injury
- Date Received
- October 13, 2020
- Date of Event
- September 23, 2020
- Report Date
- October 13, 2020
- Manufacturer
- JOHNSON & JOHNSON SURGICAL VISION, INC.
- Product Code
- OOE
- PMA / PMN Number
- K121091
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SG
- Reporter Occupation
- PHYSICIAN
Narratives
REPORTER PHONE #: (B)(6). THE SYSTEM WAS EVALUATED BY A FIELD SERVICE ENGINEER. THE BUBBLE FLATNESS WAS OUT OF SPECIFICATION AT 6.5MM TEST. THE SETTING THAT WAS UPDATED AFTER THE ISSUE WAS ZED AND ZL VALUE AND THE OPTICAL COHERENCE TOMOGRAPHY (OCT) LOOKUP TABLE WAS PERFORMED. THE ISSUE WAS RESOLVED. A RECORD REVIEW WAS PERFORMED. A PRODUCT DEFICIENCY REVIEW WAS PERFORMED AND THERE IS NO PRODUCT DEFICIENCY IDENTIFIED. A DOCUMENT, SERVICE HISTORY, AND TRENDING WAS REVIEWED. THERE IS NOT A RECOGNIZABLE ADVERSE TREND. THE RISKS AND MITIGATIONS ASSOCIATED WITH THE COMPLAINT ISSUE ARE IDENTIFIED IN EXISTING RISK DOCUMENTS AND NO NEW RISKS WERE IDENTIFIED AS PART OF THIS INVESTIGATION. A LABELING REVIEW WAS CONDUCTED; THE OPERATOR MANUAL FOR THE SYSTEM WAS REVIEWED AND FOUND TO INCLUDE ADEQUATE INSTRUCTIONS FOR USE, WARNINGS AND OPERATIONAL ERRORS. THE REVIEW OF THE DEVICE HISTORY RECORD (DHR) FOR CATALYS LASER SYSTEM SHOWED THAT THERE WERE NO ISSUES OR NON-CONFORMITIES. THE SYSTEM AND ITS COMPONENTS MET ALL SPECIFICATIONS PRIOR TO BEING RELEASED. MANUFACTURING HAS BEEN RULED OUT AS A POTENTIAL CAUSE FOR THE REPORTED ISSUE. BASED ON THE INVESTIGATION RESULTS, NO CORRECTIVE ACTION HAS BEEN ISSUED. BASED ON THE INVESTIGATION RESULTS THERE IS NO INDICATION OF A PRODUCT QUALITY DEFICIENCY. JOHNSON & JOHNSON SURGICAL VISION WILL CONTINUE TO MONITOR THIS TYPE OF COMPLAINT. ALL PERTINENT INFORMATION AVAILABLE TO JOHNSON & JOHNSON SURGICAL VISION, INC HAS BEEN SUBMITTED.
TORIC MARK CORNEA INCISION PENETRATING POSTERIOR CORNEA WAS REPORTED. A BRIEF DESCRIPTION FROM THE SURGERY CENTER INDICATED THAT DURING THE CATALYS LASER PROCEDURE ON (B)(6) 2020, THE TORIC MARK CORNEA INCISION PENETRATED THE PATIENT¿S CORNEA. THERE WAS NO ADDITIONAL PATIENT INJURY OR SURGICAL INTERVENTION REPORTED. THE SURGERY WAS COMPLETED SUCCESSFULLY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1133393 | CATALYS SYSTEM | CATALYS | OOE | JOHNSON & JOHNSON SURGICAL VISION, INC. | CATALYS-I |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |