FDA Adverse Event Malfunction Summary report: N

INRATIO

MDR report key: 1067304 · Received June 27, 2008

Report

Report Number
2954730-2008-00440
Event Type
Malfunction
Date Received
June 27, 2008
Date of Event
June 6, 2008
Report Date
June 26, 2008
Manufacturer
HEMOSENSE, INC.
Product Code
JPA
Removal / Correction Number
NI
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

INRATIO PRECISION DATA PROVIDED BY END-USER LOT 070615: PT#1 FIRST TEST INR = 4.0, SECOND TEST INR = 1.7, THIRD TEST INR = 2.4; MEAN = 2.7; SD = 1.2; %CV = 43%. THE %CV IS GREATER THAN 20%. PER INTERNAL PROCEDURE, THE PRECISION FAILED THE CRITERIA FOR PRECISION. PRODUCTS WILL BE TESTED. INRATIO PRECISION DATA PROVIDED BY END-USER LOT 070615: PT#2 FIRST TEST INR = 2.5, SECOND TEST INR = 3.2, THIRD TEST INR = 4.3; MEAN = 3.3; SD = 0.9; %CV = 27%. THE %CV IS GREATER THAN 20%. PER INTERNAL PROCEDURE, THE PRECISION FAILED THE CRITERIA FOR PRECISION. PRODUCTS WILL BE TESTED. INRATIO PRECISION DATA PROVIDED BY END-USER LOT 070615: PT#3 FIRST TEST INR = 1.7, SECOND TEST INR = 2.4; MEAN = 2.05; SD = 0.49; %CV = 24%. THE %CV IS GREATER THAN 20%. PER INTERNAL PROCEDURE, THE PRECISION FAILED THE CRITERIA FOR PRECISION. PRODUCTS WILL BE TESTED. INRATIO PRECISION DATA PROVIDED BY END-USER LOT 070615: PT# 4 FIRST TEST INR = 6.4, SECOND TEST INR = 7.0; MEAN = 6.7; SD = 0.42; %CV = 6.3%. THE %CV IS GREATER THAN 20%. PER INTERNAL PROCEDURE, THE PRECISION FAILED THE CRITERIA FOR PRECISION. PRODUCTS WILL BE TESTED.

Description of Event or Problem · 1

CALLER ALLEGED IMPRECISION WITH INRATIO METER. RESULTS AS FOLLOWS: PT#1 FIRST TEST INR = 4.0, SECOND TEST INR = 1.7, THIRD TEST INR = 2.4. PT#2 FIRST TEST INR = 2.5, SECOND TEST INR = 3.2, THIRD TEST INR = 4.3. PT#3 FIRST TEST INR = 1.7, SECOND TEST INR = 2.4. PT#4 FIRST TEST INR = 6.4, SECOND TEST INR = 7.0.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 INRATIO PROTHROMBIN TIME TEST JPA HEMOSENSE, INC. 0100004 070615

Patients

Seq Age Sex Outcome Treatment
1 NI