FDA Adverse Event Malfunction Summary report: N

BD PRECISIONGLIDE NEEDLE 30G X 1/2

MDR report key: 10665216 · Received October 12, 2020

Report

Report Number
1911916-2020-00952
Event Type
Malfunction
Date Received
October 12, 2020
Date of Event
September 1, 2020
Report Date
September 24, 2020
Manufacturer
BD MEDICAL (BD WEST) MEDICAL SURGICAL
Product Code
FMI
UDI-DI
30382903051060
PMA / PMN Number
K021475
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4). INVESTIGATION: A DEVICE HISTORY RECORD REVIEW WAS COMPLETED FOR PROVIDED MATERIAL NUMBER 305106 AND LOT NUMBER 0022949. THE REVIEW DID NOT REVEAL ANY DETECTED QUALITY ISSUES DURING THE PRODUCTION PROCESS THAT COULD HAVE CONTRIBUTED TO THIS REPORTED DEFECT. TO AID IN THE INVESTIGATION, FIVE HUNDRED PHYSICAL SAMPLES WERE RECEIVED FOR EVALUATION BY OUR QUALITY TEAM. THE SAMPLES WERE RECEIVED IN SHELF BOXES, ONE HUNDRED SAMPLES IN EACH BOX AND ALL IN THE SEALED PACKAGING BLISTERS. TWENTY RANDOM SAMPLES FROM EACH BOX, A TOTAL OF ONE HUNDRED SAMPLES, WERE PULLED FOR INVESTIGATION. EACH SAMPLE WAS CONNECTED TO A SYRINGE WITH A SALINE SOLUTION AND ALL HAD A NORMAL FLOW. BASED ON THE INVESTIGATION RESULTS, THE RANDOM SAMPLES PULLED DID NOT SHOW THE SYMPTOM REPORTED BY THE CUSTOMER AND AN EXACT CAUSE FOR THIS INCIDENT COULD NOT BE IDENTIFIED.

Description of Event or Problem · 1

IT WAS REPORTED THAT AN UNSPECIFIED NUMBER OF BD PRECISIONGLIDE¿ NEEDLE 30G X 1/2 EXPERIENCED CLOGGED/BLOCKED NEEDLES WHICH WAS NOTED PRIOR TO USE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: MATERIAL NO: 305106 BATCH NO: 0022949. EVENT DESCRIPTION: THE CUSTOMER REPORTS THAT THE NEEDLES APPEARED TO BE BLOCKED/CLOGGED OR OTHERWISE UNABLE TO PROPERLY ADMINISTER INJECTIONS. NO MEDICAL ATTENTION REQUIRED. IF ANY FURTHER INFORMATION IS OBTAINED WE WILL ADVISE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1131554 BD PRECISIONGLIDE NEEDLE 30G X 1/2 HYPODERMIC SINGLE LUMEN NEEDLE FMI BD MEDICAL (BD WEST) MEDICAL SURGICAL 305106 0022949 30382903051060

Patients

Seq Age Sex Outcome Treatment
1 Other