BD RAPID DETECTION OF SARS-COV-2 VERITOR
Report
- Report Number
- 3006948883-2020-00540
- Event Type
- Malfunction
- Date Received
- October 12, 2020
- Date of Event
- September 11, 2020
- Report Date
- March 26, 2021
- Manufacturer
- BD RAPIS DIAGNOSTICS (SUZHOU) CO. LTD.
- Product Code
- QKP
- PMA / PMN Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MO, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
EUA#: EUA201889. H6: INVESTIGATION SUMMARY BD HAS RECEIVED SEVERAL CUSTOMER COMPLAINTS FOR FALSE POSITIVE RESULTS, WHEN USING BD SARS-COV-2 REAGENTS FOR BD VERITOR¿ SYSTEM. THE CURRENT INVESTIGATION CONCERNS MULTIPLE LOTS OF BD SARS-COV-2 REAGENTS FOR BD VERITOR¿ SYSTEM. BD TAKES A SYSTEMATIC APPROACH TO INVESTIGATING FALSE POSITIVE COMPLAINTS THAT ARE RECEIVED. THIS APPROACH INVOLVES REVIEW OF MANUFACTURING BATCH HISTORY RECORDS, TESTING OF RETENTION SAMPLES, AND TESTING OF CUSTOMER RETURNED SAMPLES IF APPLICABLE. RETAIN SAMPLES WERE TESTED FOR BATCH #: 0213905. THE DEFECT COULD NOT BE REPLICATED AND THEREFORE THE COMPLAINT COULD NOT BE CONFIRMED. A DEVICE HISTORY REVIEW WAS CONDUCTED FOR LOT NUMBER 0213905. THE INVESTIGATION DID NOT FIND A ROOT CAUSE FOR THE FALSE POSITIVE RESULTS THAT WERE OBSERVED. IT IS RECOMMENDED THAT EACH CUSTOMER REVIEW THEIR WORKFLOW CAREFULLY TO ENSURE THAT THE PACKAGE INSERT IS BEING FOLLOWED AS WRITTEN. THE ROOT CAUSE IS UNDER INVESTIGATION AND WILL BE DOCUMENTED IN OUR QUALITY SYSTEM. BD CANNOT CONFIRM THE COMPLAINT BASED ON THE INVESTIGATION THAT WAS PERFORMED. A CORRECTIVE AND PREVENTIVE ACTION (CAPA#1878253) IS ALREADY INITIATED TO INVESTIGATE THE ROOT CAUSE AND SOME MITIGATION ACTIONS ARE ALREADY BEING ADDRESSED. H3 OTHER TEXT : SEE H10
IT WAS REPORTED WHILE USING BD RAPID DETECTION OF SARS-COV-2 VERITOR ASSAY FIVE (5) FALSE POSITIVE RESULTS WERE OBTAINED. REPEAT TESTS WERE PERFORMED USING BOTH VERITOR AND PCR TEST METHODS AND THE RESULTS WERE NEGATIVE. THE CUSTOMER STATED THE PATIENTS TESTED WERE ASYMPTOMATIC WITH NO KNOWN EXPOSURE. THIS TEST IS NOT INTENDED FOR USE ON ASYMPTOMATIC PATIENTS AND WAS THEREFORE USED OFF LABEL. STAFF MEMBERS WERE SENT HOME TO SELF-QUARANTINE. THERE WAS NO REPORT OF PATIENT IMPACT. EUA201889.
(B)(4). A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED. (B)(4).
IT WAS REPORTED WHILE USING BD RAPID DETECTION OF SARS-COV-2 VERITOR ASSAY FIVE (5) FALSE POSITIVE RESULTS WERE OBTAINED. REPEAT TESTS WERE PERFORMED USING BOTH VERITOR AND PCR TEST METHODS AND THE RESULTS WERE NEGATIVE. THE CUSTOMER STATED THE PATIENTS TESTED WERE ASYMPTOMATIC WITH NO KNOWN EXPOSURE. THIS TEST IS NOT INTENDED FOR USE ON ASYMPTOMATIC PATIENTS AND WAS THEREFORE USED OFF LABEL. STAFF MEMBERS WERE SENT HOME TO SELF-QUARANTINE. THERE WAS NO REPORT OF PATIENT IMPACT. (B)(4).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1131545 | BD RAPID DETECTION OF SARS-COV-2 VERITOR | CORONAVIRUS ANTIGEN DETECTION SYSTEM | QKP | BD RAPIS DIAGNOSTICS (SUZHOU) CO. LTD. | 0213905 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |