FDA Adverse Event Malfunction Summary report: N

BD RAPID DETECTION OF SARS-COV-2 VERITOR ASSAY

MDR report key: 10663248 · Received October 12, 2020

Report

Report Number
1119779-2020-00613
Event Type
Malfunction
Date Received
October 12, 2020
Date of Event
September 17, 2020
Report Date
January 14, 2021
Manufacturer
BECTON, DICKINSON & CO. (SPARKS)
Product Code
QKP
UDI-DI
00382902560821
PMA / PMN Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

H.6. INVESTIGATION: BD HAS RECEIVED SEVERAL CUSTOMER COMPLAINTS FOR FALSE POSITIVE RESULTS, WHEN USING BD SARS-COV-2 REAGENTS FOR BD VERITOR¿ SYSTEM. THE CURRENT INVESTIGATION CONCERNS MULTIPLE LOTS OF BD SARS-COV-2 REAGENTS FOR BD VERITOR¿ SYSTEM. BD TAKES A SYSTEMATIC APPROACH TO INVESTIGATING FALSE POSITIVE COMPLAINTS THAT ARE RECEIVED. THIS APPROACH INVOLVES REVIEW OF MANUFACTURING BATCH HISTORY RECORDS, TESTING OF RETENTION SAMPLES, AND TESTING OF CUSTOMER RETURNED SAMPLES IF APPLICABLE. DUE TO UNKNOWN LOT# A DHR COULD NOT BE PERFORMED. THE INVESTIGATION DID NOT FIND A ROOT CAUSE FOR THE FALSE POSITIVE RESULTS THAT WERE OBSERVED. IT IS RECOMMENDED THAT EACH CUSTOMER REVIEW THEIR WORKFLOW CAREFULLY TO ENSURE THAT THE PACKAGE INSERT IS BEING FOLLOWED AS WRITTEN. THE ROOT CAUSE IS UNDER INVESTIGATION AND WILL BE DOCUMENTED IN OUR QUALITY SYSTEM. BD CANNOT CONFIRM THE COMPLAINT BASED ON THE INVESTIGATION THAT WAS PERFORMED. A CORRECTIVE AND PREVENTIVE ACTION (CAPA#1878253) IS ALREADY INITIATED TO INVESTIGATE THE ROOT CAUSE AND SOME MITIGATION ACTIONS ARE ALREADY BEING ADDRESSED.

Description of Event or Problem · 0

BD WAS MADE AWARE OF A SURVEY THAT WAS DISTRIBUTED TO SKILLED NURSING FACILITIES USING THE VERITOR ANALYZER AND ASSAY FOR RAPID DETECTION OF SARS-COV-2. SURVEY RESULTS REVEALED THAT SOME FACILITIES HAVE POTENTIALLY RECEIVED FALSE POSITIVE RESULTS WHILE USING BD RAPID DETECTION OF SARS-COV-2 VERITOR ASSAY. SURVEY RESPONDENTS STATED THEY ARE SENDING POSITIVE SAMPLES FOR CONFIRMATION TESTING BY PCR. SOME OF THE SITES PARTICIPATING IN THE SURVEY STATED THEY ARE TESTING ASYMPTOMATIC AND SYMPTOMATIC PATIENTS AND STAFF. NOTE: THIS TEST IS NOT INTENDED FOR USE ON ASYMPTOMATIC PATIENTS. ATTEMPTS HAVE BEEN MADE TO CONTACT CUSTOMERS FOR ADDITIONAL INFORMATION, BD HAS NOT RECEIVED A RESPONSE AT THIS TIME. EUA#: (B)(4).

Additional Manufacturer Narrative · 1

MEDICAL DEVICE EXPIRATION DATE: UNKNOWN. DEVICE MANUFACTURE DATE: UNKNOWN. A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED, BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED. (B)(4).

Description of Event or Problem · 1

BD WAS MADE AWARE OF A SURVEY THAT WAS DISTRIBUTED TO SKILLED NURSING FACILITIES USING THE VERITOR ANALYZER AND ASSAY FOR RAPID DETECTION OF SARS-COV-2. SURVEY RESULTS REVEALED THAT SOME FACILITIES HAVE POTENTIALLY RECEIVED FALSE POSITIVE RESULTS WHILE USING BD RAPID DETECTION OF SARS-COV-2 VERITOR ASSAY. SURVEY RESPONDENTS STATED THEY ARE SENDING POSITIVE SAMPLES FOR CONFIRMATION TESTING BY PCR. SOME OF THE SITES PARTICIPATING IN THE SURVEY STATED THEY ARE TESTING ASYMPTOMATIC AND SYMPTOMATIC PATIENTS AND STAFF. NOTE: THIS TEST IS NOT INTENDED FOR USE ON ASYMPTOMATIC PATIENTS. ATTEMPTS HAVE BEEN MADE TO CONTACT CUSTOMERS FOR ADDITIONAL INFORMATION, BD HAS NOT RECEIVED A RESPONSE AT THIS TIME. EUA#: (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1131678 BD RAPID DETECTION OF SARS-COV-2 VERITOR ASSAY CORONAVIRUS ANTIGEN DETECTION SYSTEM QKP BECTON, DICKINSON & CO. (SPARKS) UNKNOWN 00382902560821

Patients

Seq Age Sex Outcome Treatment
1 Other