BLANK FIELDS ON THIS FORM INDICATE THE INFORMATION IS UNKNOWN, UNAVAILABLE, OR UNCHANGED. INVESTIGATION: INVESTIGATION IS REOPENED DUE TO ADDITIONAL INFORMATION PROVIDED. THE REPORTED ALLEGATIONS HAVE BEEN FURTHER INVESTIGATED BASED ON THE INFORMATION PROVIDED TO DATE. THE FOLLOWING ALLEGATIONS HAVE BEEN INVESTIGATED: VENA CAVA (VC) PERFORATION, STENOSIS, VC DILATION, FEAR, ANXIETY, PHYSICAL LIMITATIONS. FILTER INTERACTS WITH IVC WALL, E.G. PENETRATION/PERFORATION/EMBEDMENT. THIS MAY BE EITHER SYMPTOMATIC OR ASYMPTOMATIC. POTENTIAL CAUSES MAY INCLUDE IMPROPER DEPLOYMENT; AND (OR) EXCESSIVE FORCE OR MANIPULATIONS NEAR AN IN-SITU FILTER (E.G., A SURGICAL OR ENDOVASCULAR PROCEDURE IN THE VICINITY OF A FILTER). POTENTIAL ADVERSE EVENTS THAT MAY OCCUR INCLUDE, BUT ARE NOT LIMITED TO, THE FOLLOWING: TRAUMA TO ADJACENT STRUCTURES, VASCULAR TRAUMA, VENA CAVA PERFORATION, VENA CAVA PENETRATION. IVC OCCLUSION/ THROMBOSIS, NEW DVT, IVC STENOSIS AS A REPORTED COMPLICATION, IS A KNOWN RISK IN RELATION TO FILTER IMPLANT AND IS WELL DOCUMENTED IN THE CLINICAL LITERATURE AND IN CLINICAL PRACTICE GUIDELINES. THIS IS SUPPORTED BY THE CLINICAL EVIDENCE REPORT ESTABLISHED TO ASSESS AVAILABLE CLINICAL DATA TO IDENTIFY AND EVALUATE THE CLINICAL SAFETY AND PERFORMANCE OF THE COOK VENA CAVA FILTERS. POTENTIAL ADVERSE EVENTS THAT MAY OCCUR INCLUDE, BUT ARE NOT LIMITED TO, THE FOLLOWING: VENA CAVA OCCLUSION OR THROMBOSIS, VENA CAVA STENOSIS, DEEP VEIN THROMBOSIS. UNKNOWN IF THE REPORTED VENA CAVA DILATION, FEAR, ANXIETY, AND PHYSICAL LIMITATIONS ARE DIRECTLY RELATED TO THE FILTER AND UNABLE TO IDENTIFY A CORRESPONDING FAILURE MODE AT THIS POINT IN TIME. CATALOG NUMBER AND LOT NUMBER ARE UNKNOWN. THE PRODUCT IS MANUFACTURED AND INSPECTED ACCORDING TO SPECIFICATIONS. NO EVIDENCE TO SUGGEST THAT THIS DEVICE WAS NOT MANUFACTURED ACCORDING TO SPECIFICATIONS AND NOTHING INDICATES THAT THE FILTER DID NOT PERFORM AS INTENDED, E.G. INTENDED FOR THE PREVENTION OF RECURRENT PULMONARY EMBOLISM (PE) VIA PLACEMENT IN THE VENA CAVA. COOK WILL REOPEN ITS INVESTIGATION IF FURTHER INFORMATION IS RECEIVED WARRANTING SUPPLEMENTATION IN ACCORDANCE WITH 21 C.F.R. 803.56. THIS REPORT INCLUDES INFORMATION KNOWN AT THIS TIME. A FOLLOW-UP MEDWATCH REPORT WILL BE SUBMITTED IF ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE. THIS REPORT IS REQUIRED BY THE FDA UNDER 21 CFR PART 803. THIS REPORT IS BASED ON UNCONFIRMED INFORMATION SUBMITTED BY OTHERS. NEITHER THE SUBMISSION OF THIS REPORT NOR ANY STATEMENT MADE IN IT IS INTENDED TO BE AN ADMISSION THAT ANY COOK DEVICE IS DEFECTIVE OR MALFUNCTIONED, THAT A DEATH OR SERIOUS INJURY OCCURRED, OR THAT ANY COOK DEVICE CAUSED OR CONTRIBUTED TO, OR IS LIKELY TO CAUSE OR CONTRIBUTE TO A DEATH OR SERIOUS INJURY IF A MALFUNCTION OCCURRED.
PATIENT ALLEGEDLY RECEIVED AN IMPLANT ON (B)(6) 2007 VIA THE RIGHT COMMON FEMORAL VEIN DUE TO MOTOR VEHICLE ACCIDENT (MVA) WITH MULTIPLE FRACTURES. PATIENT IS ALLEGING VENA CAVA/ORGAN PERFORATION. THE PATIENT FURTHER ALLEGES FEAR, ANXIETY, PHYSICAL LIMITATIONS. REPORT FROM CT (COMPUTED TOMOGRAPHY): "THE IVC FILTER IS IDENTIFIED. THE TIP APPEARS TO AT THE LEVEL OF THE RENAL VEINS. I BELIEVE THE TIP IS ALSO AT THE ANTERIOR MARGIN OF THE INFERIOR VENA CAVA. 4 PRONGS EXTEND PERIPHERAL TO THE VENA CAVA; ONE IS AT THE SEROSAL MARGIN OF THE DUODENUM. THE ILIAC VESSELS APPEAR RELATIVELY PROMINENT WITH A DISTAL IVC SOMEWHAT DIMINISHED IN DIAMETER AND WITH THE MORE PROXIMAL IVC INCREASED SIZE. ABNORMALLY THERE ARE VARICOSITIES IN THE RIGHT HEMIABDOMEN."
OCCUPATION: NON-HEALTHCARE PROFESSIONAL. INVESTIGATION: IT HAS NOT BEEN POSSIBLE TO FURTHER INVESTIGATE OR EVALUATE THIS ALLEGED EVENT BASED ON THE LIMITED INFORMATION AND/OR NO DEVICE FAILURE PROVIDED TO DATE. CATALOG NUMBER AND LOT NUMBER ARE UNKNOWN, HOWEVER, THE ALLEGED TULIP IS MANUFACTURED AND INSPECTED ACCORDING TO SPECIFICATIONS. NO EVIDENCE TO SUGGEST THAT THIS DEVICE WAS NOT MANUFACTURED ACCORDING TO SPECIFICATIONS AND NOTHING INDICATES THAT THE FILTER DID NOT PERFORM AS INTENDED, E.G. INTENDED FOR THE PREVENTION OF RECURRENT PULMONARY EMBOLISM (PE) VIA PLACEMENT IN THE VENA CAVA. COOK WILL REOPEN ITS INVESTIGATION IF FURTHER INFORMATION IS RECEIVED WARRANTING SUPPLEMENTATION IN ACCORDANCE WITH 21 C.F.R. 803.56. THIS REPORT INCLUDES INFORMATION KNOWN AT THIS TIME. A FOLLOW-UP MEDWATCH REPORT WILL BE SUBMITTED IF ADDITIONAL RELEVANT INFORMATION BECOMES AVAILABLE.
IT IS ALLEGED THAT THE PATIENT RECEIVED A GUNTHER TULIP ON (B)(6) 2007. IT IS ALLEGED THAT THE PATIENT WAS INJURED WITHOUT FURTHER EXPLANATION. HOSPITAL AND MEDICAL RECORDS HAVE BEEN REQUESTED, BUT NOT YET PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1127050 | UNKNOWN | DTK FILTER, INTRAVASCULAR, CARDIOVASCULAR | DTK | COOK INC | UNKNOWN |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |