FDA Adverse Event Malfunction Summary report: N

CELLEX PHOTOPHERESIS SYSTEM

MDR report key: 10657771 · Received October 9, 2020

Report

Report Number
2523595-2020-00108
Event Type
Malfunction
Date Received
October 9, 2020
Date of Event
September 9, 2020
Report Date
November 6, 2020
Manufacturer
THERAKOS, INC.
Product Code
LNR
PMA / PMN Number
P860003
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 0

THE COMPLAINT KIT WITH SMART CARD WAS RETURNED FOR EVALUATION. THE SMART CARD DATA SHOWS THE TREATMENT PROCEEDED THROUGH THE PHOTOACTIVATION PHASE OF THE PROCEDURE UNTIL THE OPERATOR ABORTED THE TREATMENT. THE PUMP TUBING ORGANIZER (PTO) WAS THE ONLY COMPONENT OF THE KIT RETURNED FOR ANALYSIS. EXAMINATION OF THE RETURNED PTO FOUND DRIED BLOOD ON THE RIGHT SIDE OF THE BLOOD FILTER IN THE PTO. THE PTO WAS PRESSURE TESTED TO CHECK FOR LEAKS AND NO LEAKS WERE IDENTIFIED. A MATERIAL TRACE OF THE RELATED COMPONENTS USED TO BUILD LOT J108 SHOWED NO RELATED NON-CONFORMANCES. THE DEVICE HISTORY RECORD (DHR) REVIEW DID NOT RESULT IN ANY RELATED NON-CONFORMANCES AND THIS KIT LOT PASSED ALL LOT RELEASE TESTING. THE PTO LEAK WAS VERIFIED BASED ON THE DRIED BLOOD FOUND ON THE INSIDE OF THE PTO; HOWEVER, PRESSURE TESTING OF THE RETURNED PTO DID NOT RESULT IN A LEAK. THE CAUSE OF THE PTO LEAK WAS MOST LIKELY A WEAK WELD AT THE BLOOD FILTER IN THE PTO. THE ROOT CAUSE FOR THE WEAK WELD COULD NOT BE DETERMINED BASED ON THE AVAILABLE INFORMATION. NO FURTHER ACTION IS REQUIRED AT THIS TIME. THIS INVESTIGATION IS NOW COMPLETE. (B)(4). P.T. (B)(6) 2020.

Additional Manufacturer Narrative · 1

THE SYSTEM WAS USED FOR TREATMENT. THIS CASE IS REPORTABLE AS A MDR DUE TO THE REPORTABLE MALFUNCTION PUMP TUBING ORGANIZER (PTO) LEAK. SINCE THIS REPORTABLE MALFUNCTION IS ONLY ASSOCIATED WITH THE KIT, THIS MDR WILL ONLY BE AGAINST THE KIT. A BATCH RECORD REVIEW FOR KIT LOT J108 WAS CONDUCTED. THERE WERE NO NON-CONFORMANCES ASSOCIATED WITH THIS LOT. THIS LOT MET ALL RELEASE REQUIREMENTS. A REVIEW OF KIT LOT J108 SHOWS NO TRENDS. TRENDS WERE REVIEWED FOR COMPLAINT CATEGORY, PTO LEAK. NO TRENDS WERE DETECTED FOR THIS COMPLAINT CATEGORY. THIS ASSESSMENT IS BASED ON THE INFORMATION AVAILABLE AT THE TIME OF THE INVESTIGATION. AT THE TIME OF THIS REPORT, THE ANALYSIS OF THE RETURNED KIT AND PHOTOGRAPH IS STILL IN PROGRESS. A SUPPLEMENTAL REPORT WILL BE FILED WHEN THE ANALYSIS IS COMPLETE. (B)(4).

Description of Event or Problem · 1

THE CUSTOMER CONTACTED MALLINCKRODT TO REPORT THEY EXPERIENCED A PUMP TUBING ORGANIZER (PTO) LEAK WITH THEIR CELLEX PHOTOPHERESIS KIT ("KIT") DURING AN EXTRACORPOREAL PHOTOPHERESIS (ECP) TREATMENT. THE CUSTOMER REPORTED A BLOOD LEAK WAS OBSERVED IN THE PUMP TUBING ORGANIZER UNDERNEATH THE BLOOD FILTER DURING THE PHOTOACTIVATION PHASE OF THE PROCEDURE. THE CUSTOMER ABORTED THE ECP TREATMENT AND DID NOT RETURN RESIDUAL BLOOD WITHIN THE KIT TO THE PATIENT. THE CUSTOMER RETURNED THE KIT AND A PHOTOGRAPH FOR INVESTIGATION.