BACT/ALERT MP CULTURE BOTTLES
Report
- Report Number
- 3002769706-2008-00002
- Event Type
- Malfunction
- Date Received
- June 25, 2008
- Date of Event
- May 27, 2008
- Report Date
- June 20, 2008
- Manufacturer
- BIOMERIEUX, INC.
- Product Code
- JTA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UK
- Reporter Occupation
- UNKNOWN
Narratives
THE USER HAD DISINFECTED THE PLASTIC BACT/ALERT MP CULTURE BOTTLE WITH BIOCLEANSE (TEKNOM), A DISINFECTANT, WHICH ACCORDING TO TEKNON'S APPLICATION CHART, IS NOT RECOMMENDED FOR THE USE ON POLYCARBONATE BOTTLES. BIOMERIEUX, INC. HAS INVESTIGATED THE INTERACTION OF SOME DISINFECTANTS ON ITS POLYCARBONATE BOTTLES AND HAS ISSUED A NOTICE ALERTING USERS TO READ THE DISINFECTANT'S LABEL TO DETERMINE COMPATIBILITY WITH POLYCARBONATE BOTTLES. BIOMERIEUX HAS ALSO MODIFIED ITS PACKAGE INSERT WITH A CAUTION ABOUT POLYCARBONATES AND CHOICE OF DISINFECTANTS BY THE USER.
A PLASTIC BACT/ALERT MP CULTURE BOTTLE WAS FOUND BROKEN IN THE BACT/ALERT INCUBATOR. AFB STAIN INDICATED THAT ITS CONTENTS WERE NEGATIVE. THERE WERE NO ADVERSE EVENTS NOTED AS A RESULT OF THIS INCIDENT. THE LAB WAS CLOSED AND A FORMALDEHYDE GAS STERILIZATION WAS PERFORMED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | BACT/ALERT MP CULTURE BOTTLES | MICROBIAL GROWTH MONITOR | JTA | BIOMERIEUX, INC. | 1018685 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |