FDA Adverse Event Malfunction Summary report: N

BACT/ALERT MP CULTURE BOTTLES

MDR report key: 1065518 · Received June 25, 2008

Report

Report Number
3002769706-2008-00002
Event Type
Malfunction
Date Received
June 25, 2008
Date of Event
May 27, 2008
Report Date
June 20, 2008
Manufacturer
BIOMERIEUX, INC.
Product Code
JTA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THE USER HAD DISINFECTED THE PLASTIC BACT/ALERT MP CULTURE BOTTLE WITH BIOCLEANSE (TEKNOM), A DISINFECTANT, WHICH ACCORDING TO TEKNON'S APPLICATION CHART, IS NOT RECOMMENDED FOR THE USE ON POLYCARBONATE BOTTLES. BIOMERIEUX, INC. HAS INVESTIGATED THE INTERACTION OF SOME DISINFECTANTS ON ITS POLYCARBONATE BOTTLES AND HAS ISSUED A NOTICE ALERTING USERS TO READ THE DISINFECTANT'S LABEL TO DETERMINE COMPATIBILITY WITH POLYCARBONATE BOTTLES. BIOMERIEUX HAS ALSO MODIFIED ITS PACKAGE INSERT WITH A CAUTION ABOUT POLYCARBONATES AND CHOICE OF DISINFECTANTS BY THE USER.

Description of Event or Problem · 1

A PLASTIC BACT/ALERT MP CULTURE BOTTLE WAS FOUND BROKEN IN THE BACT/ALERT INCUBATOR. AFB STAIN INDICATED THAT ITS CONTENTS WERE NEGATIVE. THERE WERE NO ADVERSE EVENTS NOTED AS A RESULT OF THIS INCIDENT. THE LAB WAS CLOSED AND A FORMALDEHYDE GAS STERILIZATION WAS PERFORMED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 BACT/ALERT MP CULTURE BOTTLES MICROBIAL GROWTH MONITOR JTA BIOMERIEUX, INC. 1018685

Patients

Seq Age Sex Outcome Treatment
1