FDA Adverse Event Injury Summary report: N

IMPLANT:SPLN HA THD 13MM

MDR report key: 10654388 · Received October 9, 2020

Report

Report Number
0002023141-2020-01604
Event Type
Injury
Date Received
October 9, 2020
Date of Event
March 24, 2019
Report Date
March 16, 2021
Manufacturer
ZIMMER DENTAL
Product Code
DZE
PMA / PMN Number
K013494
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
IS
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

ZIMMER BIOMET COMPLAINT (B)(4).

Description of Event or Problem · 0

NO FURTHER EVENT INFORMATION IS AVAILABLE AT THE TIME OF THIS REPORT.

Additional Manufacturer Narrative · 1

ZIMMER BIOMET (B)(4). WEIGHT UNKNOWN / NOT PROVIDED. DATE OF EVENT UNKNOWN / NOT PROVIDED. EXPIRATION DATE AND UDI UNKNOWN. PMA/510(K) NUMBER: K943604. DEVICE MANUFACTURE DATE UNKNOWN. A SUMMARY INVESTIGATION HAS BEEN COMPLETED FOR PERI-IMPLANTITIS EVENTS RECOGNIZING THAT A DEFINITIVE ROOT CAUSE CANNOT BE IDENTIFIED DUE TO A WIDE RANGE OF EXTERNAL FACTORS (NON-DESIGN OR MANUFACTURING RELATED), INCLUDING MEDICAL CONDITIONS (E.G., DIABETES, POOR BONE QUALITY, ETC.) / PATIENT HABITS (E.G., SMOKING) AND SURGICAL TECHNIQUE. PREVIOUSLY COMPLETED INVESTIGATIONS FOR THESE EVENTS HAVE NOT IDENTIFIED ANY SIGNALS INDICATING POTENTIAL NON-CONFORMANCES AFFECTING THE MANUFACTURING AND STERILIZATION PROCESSES. FURTHERMORE, THE PROBABILITY OF A MANUFACTURING OR DESIGN DEFECT THAT MIGHT LEAD TO PERI-IMPLANTITIS OCCURRING AND ESCAPING THE AVAILABLE DETECTIONS HAS BEEN ASSESSED AND FOUND REMOTE AND ALMOST NONEXISTENT. SHOULD ADDITIONAL INFORMATION BE RECEIVED WHICH INDICATES THAT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT, AN ADDITIONAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE IMPLANT WAS REMOVED DUE TO PERI-IMPLANTITIS AND INFLAMMATION. IMPLANT HAD BEEN RESTORED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1119606 IMPLANT:SPLN HA THD 13MM DENTAL IMPLANT DZE ZIMMER DENTAL 982024

Patients

Seq Age Sex Outcome Treatment
1 52 YR Required Intervention